Zilles K, Schleicher A, Rath M, Glaser T, Traber J
J Neurosci Methods. 1986 Oct;18(1-2):207-20. doi: 10.1016/0165-0270(86)90120-2.
A computerized image processing system for quantitative receptor autoradiography on tritium-sensitive film is described. The method implements an image analyzer (IBAS 1 + 2) for the acquisition of an image which is evaluated with respect to the regional distribution of binding site densities by a small computer which is connected to a digitizer. Several methodological problems must be considered when using tritium-labelled compounds and an image analyzer. The use of tissue standards and shading procedures, the improvement of signal-to-noise ratio, the calibration of the system which establishes the non-linear relation between grey values and concentration of radioactivity, the contrast enhancement and the quenching problem are discussed. This method permits a quick and precise quantitative analysis of binding site distribution in receptor autoradiography.