Suppr超能文献

基于短扫描 CT 数据的部分角度重建的冠状动脉区域的运动补偿。

Motion compensation in the region of the coronary arteries based on partial angle reconstructions from short-scan CT data.

机构信息

Siemens Healthcare GmbH, Forchheim, Germany.

German Cancer Research Center, Heidelberg, Germany.

出版信息

Med Phys. 2017 Nov;44(11):5795-5813. doi: 10.1002/mp.12514. Epub 2017 Sep 22.

Abstract

PURPOSE

In order to mitigate motion-induced artifacts, several motion compensation (MoCo) methods have been developed, which are either able to (a) compensate for severe artifacts, but utilize the data for the reconstruction of several cardiac phases, or (b) improve image quality of a single reconstruction with only moderate motion artifacts. We propose a method combining both benefits: dose efficiency by utilizing only the data needed for a single short-scan reconstruction while still being able to compensate for severe artifacts.

METHODS

We introduce a MoCo method, which we call PAMoCo, to improve the visualization of the coronary arteries of a standard coronary CT angiography exam by reducing motion artifacts. As a first step, we segment a region of interest covering a chosen coronary artery. We subdivide a volume covering the whole heart into several stacks, which are sub-volumes, reconstructed from phase-correlated short-scan data acquired during different heart cycles. Motion-compensated reconstruction is performed for each stack separately, based on partial angle reconstructions, which are derived by dividing the data corresponding to the segmented stack volume into several double-overlapping sectors. We model motion along the coronary artery center line obtained from segmentation and the temporal dimension by a low-degree polynomial and create a dense 3D motion vector field (MVF). The parameters defining the MVF are estimated by optimizing an image artifact measuring cost function and we employ a semi-global optimization routine by re-initializing the optimization multiple times. The algorithm was evaluated on the basis of a phantom measurement and clinical data. For the phantom measurement an artificial vessel equipped with calcified lesions mounted on a moving robot arm was measured, where typical coronary artery motion patterns for 70 bpm and 90 bpm have been applied. For analysis, we calculated the calcified volume V inside an ROI and measured the maximum vessel diameter d based on cross-sectional views to compare the performances of standard reconstructions obtained via filtered backprojection (FBP) and PAMoCo reconstructions between 20% and 80% of the cardiac cycle. Further, the new algorithm was applied to six clinical cases of patients with heart rates between 50 bpm and 74 bpm. Standard FBP, PAMoCo reconstructions were performed and compared to best phase FBP reconstructions and another MoCo algorithm, which is based on motion artifact metrics (MAM), via visual inspection.

RESULTS

In case of the phantom measurement we found a strong dependence of V and d on the cardiac phase in case of the FBP reconstructions. When applying PAMoCo, V and d became almost constant due to a better discrimination from calcium to vessel and water background and values close to the ground truth have been derived. In the clinical study we chose reconstructions showing strong motion artifacts and observed a substantially improved delineation of the coronary arteries in PAMoCo reconstructions compared to the standard FBP reconstructions and also the MAM images, confirming the findings of the phantom measurement.

CONCLUSIONS

Due to the fast reconstruction of PAMoCo images and the introduction of a new motion model, we were able to re-initialize the optimization routine at pre-selected parameter sets and thereby increase the potential of the MAM algorithm. From the phantom measurement we conclude that PAMoCo performed almost equally well in all cardiac phases and suggest applying the PAMoCo algorithm for single source systems in case of patients with high or irregular heart rates.

摘要

目的

为了减轻运动引起的伪影,已经开发了几种运动补偿(MoCo)方法,这些方法要么能够(a)补偿严重的伪影,但利用重建几个心动周期的数据,要么(b)仅用中度运动伪影改善单个重建的图像质量。我们提出了一种结合两者优点的方法:利用单次短扫描重建所需的数据,同时仍能补偿严重的伪影,从而提高剂量效率。

方法

我们引入了一种 MoCo 方法,称为 PAMoCo,通过减少运动伪影来改善标准冠状动脉 CT 血管造影检查中冠状动脉的可视化。作为第一步,我们分割一个包含所选冠状动脉的感兴趣区域。我们将覆盖整个心脏的体积细分为几个堆栈,这些堆栈是从不同心动周期采集的相位相关短扫描数据重建的子体积。根据从分割堆栈体积中分割的数据,我们为每个堆栈分别进行基于部分角度重建的运动补偿重建,这些数据被分割成几个双重叠扇区。我们通过低阶多项式对沿冠状动脉中心线和时间维度的运动进行建模,并创建密集的 3D 运动向量场(MVF)。通过优化图像伪影测量成本函数来估计定义 MVF 的参数,我们通过多次重新初始化优化来使用半全局优化例程。该算法基于幻影测量和临床数据进行了评估。对于幻影测量,我们测量了安装在移动机械臂上的带有钙化病变的人造血管,其中应用了 70 bpm 和 90 bpm 的典型冠状动脉运动模式。为了分析,我们在 ROI 内计算了钙化体积 V,并根据横截面视图测量了最大血管直径 d,以比较通过滤波反投影(FBP)获得的标准重建和 PAMoCo 重建之间 20%到 80%的心脏周期的性能。此外,我们将新算法应用于 6 例心率在 50 bpm 到 74 bpm 之间的患者的临床病例。进行了标准 FBP、PAMoCo 重建,并通过视觉检查与最佳相位 FBP 重建和另一种基于运动伪影指标(MAM)的 MoCo 算法进行比较。

结果

在幻影测量中,我们发现 FBP 重建中 V 和 d 对心脏相位有很强的依赖性。当应用 PAMoCo 时,由于更好地区分钙与血管和水背景,V 和 d 变得几乎恒定,并且得出了接近真实值的结果。在临床研究中,我们选择了显示强烈运动伪影的重建,并观察到 PAMoCo 重建中冠状动脉的描绘明显优于标准 FBP 重建和 MAM 图像,这证实了幻影测量的结果。

结论

由于 PAMoCo 图像的快速重建和新运动模型的引入,我们能够在预选参数集处重新初始化优化例程,从而提高 MAM 算法的潜力。从幻影测量中我们得出结论,PAMoCo 在所有心动周期中表现几乎相同,并建议在心率高或不规则的患者中使用 PAMoCo 算法用于单源系统。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验