Propex II、Root ZX II以及影像学测量在确定主要孔位置方面的准确性评估。

assessment of accuracy of Propex II, Root ZX II, and radiographic measurements for location of the major foramen.

作者信息

Tampelini Fernanda Garcia, Coelho Marcelo Santos, Rios Marcos de Azevêdo, Fontana Carlos Eduardo, Rocha Daniel Guimarães Pedro, Pinheiro Sergio Luiz, Bueno Carlos Eduardo da Silveira

机构信息

Department of Endodontics, São Leopoldo Mandic Dental School, Campinas, SP, Brazil.

Department of Endodontics, Universidade Paulista School of Dentistry, Sorocaba, SP, Brazil.

出版信息

Restor Dent Endod. 2017 Aug;42(3):200-205. doi: 10.5395/rde.2017.42.3.200. Epub 2017 May 16.

Abstract

OBJECTIVES

The aim of this study was to assess the accuracy of 2 third-generation electronic apex locators (EALs), Propex II (Dentsply Maillefer) and Root ZX II (J. Morita), and radiographic technique for locating the major foramen (MF).

MATERIALS AND METHODS

Thirty-two premolars with single canals that required extraction were included. Following anesthesia, access, and initial canal preparation with size 10 and 15 K-flex files and SX and S1 rotary ProTaper files, the canals were irrigated with 2.5% sodium hypochlorite. The length of the root canal was verified 3 times for each tooth using the 2 apex locators and once using the radiographic technique. Teeth were extracted and the actual WL was determined using size 15 K-files under a × 25 magnification. The Biostat 4.0 program (AnalystSoft Inc.) was used for comparing the direct measurements with those obtained using radiographic technique and the apex locators. Pearson's correlation analysis and analysis of variance (ANOVA) were used for statistical analyses.

RESULTS

The measurements obtained using the visual method exhibited the strongest correlation with Root ZX II ( = 0.94), followed by Propex II ( = 0.90) and Ingle's technique ( = 0.81; < 0.001). Descriptive statistics using ANOVA (Tukey's test) revealed significant differences between the radiographic measurements and both EALs measurements ( < 0.05).

CONCLUSIONS

Both EALs presented similar accuracy that was higher than that of the radiographic measurements obtained with Ingle's technique. Our results suggest that the use of these EALs for MF location is more accurate than the use of radiographic measurements.

摘要

目的

本研究旨在评估两款第三代电子根尖定位仪(EAL)——Propex II(登士柏迈福利公司)和Root ZX II(日本森田公司)以及影像学技术在定位根尖孔(MF)方面的准确性。

材料与方法

纳入32颗需要拔除的单根管前磨牙。麻醉、开髓后,先用10号和15号K-flex锉以及SX和S1号旋转ProTaper锉进行初始根管预备,然后用2.5%次氯酸钠冲洗根管。每颗牙齿使用这两款根尖定位仪各测量3次根管长度,并用影像学技术测量1次。牙齿拔除后,在25倍放大倍数下用15号K锉确定实际工作长度(WL)。使用Biostat 4.0程序(AnalystSoft公司)将直接测量值与影像学技术及根尖定位仪获得的值进行比较。采用Pearson相关分析和方差分析(ANOVA)进行统计分析。

结果

视觉法获得的测量值与Root ZX II的相关性最强(r = 0.94),其次是Propex II(r = 0.90)和英格尔技术(r = 0.81;P < 0.001)。使用ANOVA(Tukey检验)进行描述性统计显示,影像学测量值与两款EAL测量值之间存在显著差异(P < 0.05)。

结论

两款EAL的准确性相似,均高于英格尔技术获得的影像学测量值。我们的结果表明,使用这些EAL定位MF比使用影像学测量更准确。

相似文献

5
A comparison between two electronic apex locators: an ex vivo investigation.两种电子根尖定位仪的比较:一项体外研究。
Int Endod J. 2007 May;40(5):362-73. doi: 10.1111/j.1365-2591.2006.01229.x. Epub 2007 Feb 19.

本文引用的文献

5
Clinical efficacy of electronic apex locators: systematic review.电子根尖定位仪的临床疗效:系统评价
J Endod. 2014 Jun;40(6):759-77. doi: 10.1016/j.joen.2014.03.011. Epub 2014 May 2.
10
In vitro comparison of three electronic apex locators.三种电子根尖定位仪的体外比较。
J Endod. 2010 Feb;36(2):279-81. doi: 10.1016/j.joen.2009.09.016. Epub 2009 Nov 10.