电子根尖定位仪的临床疗效:系统评价

Clinical efficacy of electronic apex locators: systematic review.

作者信息

Martins Jorge N R, Marques Duarte, Mata António, Caramês João

机构信息

School of Dental Medicine, Lisbon University, Lisbon, Portugal; Inter PG Endodontics, New York University College of Dentistry, New York, New York; Department of Endodontics, Implantology Institute, Lisbon, Portugal.

School of Dental Medicine, Lisbon University, Lisbon, Portugal; Oral Biology and Biochemistry Group, Biomedical and Oral Sciences Research Unit (FCT Unit 4062), Evidence Based Dentistry Center, Lisbon University, Lisbon, Portugal; Implantology Institute, Lisbon, Portugal.

出版信息

J Endod. 2014 Jun;40(6):759-77. doi: 10.1016/j.joen.2014.03.011. Epub 2014 May 2.

Abstract

INTRODUCTION

Apical constriction has been proposed as the most appropriate apical limit for the endodontic working length. Despite being the most used, some limitations are attributed to the radiographic method of working length determination. It lacks precision because it is based on the average position of the apical constriction. The electronic apex locators have been presented as an alternative to the odontometry performed by radiography. These devices detect the transition of the pulp to the periodontal tissue, which is anatomically very close to the apical constriction and may perform with improved accuracy.

METHODS

A systematic review was performed to compare the radiographic and electronic methods. Clinical studies that compared both methods were searched for on 7 electronic databases, a manual search was performed on the bibliography of articles collected on the electronic databases, and the authors were contacted to ask for references of more research not detected on the electronic and manual search.

RESULTS

Twenty-one articles were selected. The majority were comparative or evaluation studies, and very few clinical studies comparing both methods are available. Several methodological limitations are present on the collected articles and debated in this review.

CONCLUSIONS

Although the available scientific evidence base is short and at considerable risk of bias, it is still possible to conclude that the apical locator reduces the patient radiation exposure and also that the electronic method may perform better on the working length determination. At least one radiographic control should be performed to detect possible errors of the electronic devices.

摘要

引言

根尖缩窄被认为是根管治疗工作长度最合适的根尖界限。尽管它是最常用的,但工作长度确定的影像学方法存在一些局限性。它缺乏精确性,因为它基于根尖缩窄的平均位置。电子根尖定位仪已被提出作为影像学测量牙长度的替代方法。这些设备检测牙髓向牙周组织的转变,这在解剖学上非常接近根尖缩窄,并且可能具有更高的准确性。

方法

进行了一项系统评价以比较影像学和电子方法。在7个电子数据库中搜索比较这两种方法的临床研究,对电子数据库中收集的文章的参考文献进行手动搜索,并联系作者询问在电子和手动搜索中未发现的更多研究的参考文献。

结果

选择了21篇文章。大多数是比较性或评估性研究,很少有比较这两种方法的临床研究。收集的文章存在一些方法学上的局限性,并在本综述中进行了讨论。

结论

尽管现有的科学证据基础薄弱且存在相当大的偏倚风险,但仍可以得出结论,根尖定位仪减少了患者的辐射暴露,并且电子方法在工作长度确定方面可能表现更好。应至少进行一次影像学对照以检测电子设备可能存在的误差。