Sobral School of Dentistry, Federal University of Ceará, Fortaleza, CE, Brazil.
J Appl Oral Sci. 2013 Mar-Apr;21(2):132-7. doi: 10.1590/1678-7757201302188.
The aim of this study was to evaluate, ex vivo, the precision of five electronic root canal length measurement devices (ERCLMDs) with different operating systems: the Root ZX, Mini Apex Locator, Propex II, iPex, and RomiApex A-15, and the possible influence of the positioning of the instrument tips short of the apical foramen.
Forty-two mandibular bicuspids had their real canal lengths (RL) previously determined. Electronic measurements were performed 1.0 mm short of the apical foramen (-1.0), followed by measurements at the apical foramen (0.0). The data resulting from the comparison of the ERCLMD measurements and the RL were evaluated by the Wilcoxon and Friedman tests at a significance level of 5%.
Considering the measurements performed at 0.0 and -1.0, the precision rates for the ERCLMDs were: 73.5% and 47.1% (Root ZX), 73.5% and 55.9% (Mini Apex Locator), 67.6% and 41.1% (Propex II), 61.7% and 44.1% (iPex), and 79.4% and 44.1% (RomiApex A-15), respectively, considering ±0.5 mm of tolerance. Regarding the mean discrepancies, no differences were observed at 0.0; however, in the measurements at -1.0, the iPex, a multi-frequency ERCLMD, had significantly more discrepant readings short of the apical foramen than the other devices, except for the Propex II, which had intermediate results. When the ERCLMDs measurements at -1.0 were compared with those at 0.0, the Propex II, iPex and RomiApex A-15 presented significantly higher discrepancies in their readings.
Under the conditions of the present study, all the ERCLMDs provided acceptable measurements at the 0.0 position. However, at the -1.0 position, the ERCLMDs had a lower precision, with statistically significant differences for the Propex II, iPex, and RomiApex A-15.
本研究旨在评估 5 种不同操作系统的电子根管长度测量仪(ERCLMD)的精确性:Root ZX、Mini Apex Locator、Propex II、iPex 和 RomiApex A-15,并研究器械尖端短于根尖孔时的定位对测量结果的可能影响。
42 颗下颌双尖牙预先确定了其真实根管长度(RL)。在距离根尖孔 1.0mm 处(-1.0)进行电子测量,然后在根尖孔处(0.0)进行测量。通过 Wilcoxon 和 Friedman 检验评估 ERCLMD 测量值与 RL 之间的差异,检验水准为 5%。
考虑到 0.0 和-1.0 处的测量值,ERCLMD 的精度率分别为:73.5%和 47.1%(Root ZX)、73.5%和 55.9%(Mini Apex Locator)、67.6%和 41.1%(Propex II)、61.7%和 44.1%(iPex)和 79.4%和 44.1%(RomiApex A-15),均考虑了±0.5mm 的允许误差。关于平均差异,在 0.0 时无差异;然而,在-1.0 时的测量中,iPex(一种多频 ERCLMD)与其他器械相比,在距离根尖孔不足 1mm 处的读数差异明显更大,而 Propex II 的结果则处于中间水平。当比较 ERCLMD 在-1.0 时的测量值与在 0.0 时的测量值时,Propex II、iPex 和 RomiApex A-15 的读数差异明显更大。
在本研究条件下,所有 ERCLMD 在 0.0 位置的测量结果都可接受。然而,在-1.0 位置时,ERCLMD 的精度较低,Propex II、iPex 和 RomiApex A-15 的差异有统计学意义。