IEEE Trans Ultrason Ferroelectr Freq Control. 2017 Sep;64(9):1298-1304. doi: 10.1109/TUFFC.2017.2707981.
A lung ultrasound surface wave elastography (LUSWE) technique is developed to measure superficial lung tissue elastic properties. The purpose of this paper was to translate LUSWE into clinical studies for assessing patients with interstitial lung disease (ILD) and present the pilot data from lung measurements on 10 healthy subjects and 10 patients with ILD. ILD includes multiple lung disorders in which the lung tissue is distorted and stiffened by tissue fibrosis. Chest radiography and computed tomography are the most commonly used techniques for assessing lung disease, but they are associated with radiation and cannot directly measure lung elastic properties. LUSWE provides a noninvasive and nonionizing technique to measure the elastic properties of superficial lung tissue. LUSWE was used to measure regions of both lungs through six intercostal spaces for patients and healthy subjects. The data are presented as wave speed at 100, 150, and 200 Hz at the six intercostal spaces. As an example, the surface wave speeds are, respectively, 1.88 ± 0.11 m/s at 100 Hz, 2.74 ± 0.26 m/s at 150 Hz, and 3.62 ± 0.13 m/s at 200 Hz for a healthy subject in the upper right lung; this is in comparison to measurements from an ILD patient of 3.3 ± 0.37 m/s at 100 Hz, 4.38 ± 0.33 m/s at 150 Hz, and 5.24 ± 0.44 m/s at 200 Hz in the same lung space. Significant differences in wave speed between healthy subjects and ILD patients were found. LUSWE is a safe and noninvasive technique which may be useful for assessing ILD.
肺超声表面波弹性成像(LUSWE)技术用于测量浅层肺组织的弹性特性。本文的目的是将 LUSWE 转化为评估间质性肺病(ILD)患者的临床研究,并介绍对 10 名健康受试者和 10 名 ILD 患者进行肺测量的初步数据。ILD 包括多种肺部疾病,其中肺组织由于组织纤维化而变形和变硬。胸部 X 线摄影和计算机断层扫描是最常用于评估肺部疾病的技术,但它们与辐射有关,并且不能直接测量肺弹性特性。LUSWE 提供了一种非侵入性和非电离的技术来测量浅层肺组织的弹性特性。LUSWE 用于通过六个肋间空间测量患者和健康受试者的左右肺区。数据以六个肋间空间的 100、150 和 200Hz 的波速表示。例如,健康受试者右上肺的表面波速度分别为 100Hz 时为 1.88±0.11m/s,150Hz 时为 2.74±0.26m/s,200Hz 时为 3.62±0.13m/s;相比之下,ILD 患者同一肺区的测量值分别为 100Hz 时为 3.3±0.37m/s,150Hz 时为 4.38±0.33m/s,200Hz 时为 5.24±0.44m/s。在健康受试者和 ILD 患者之间发现了波速的显著差异。LUSWE 是一种安全且非侵入性的技术,可能有助于评估 ILD。