Assessments of range of motion (ROM) represent an essential part of clinical diagnostics. Ultrasonic movement analyses have been demonstrated to provide reliable results when analyzing complete amplitudes (e.g., flexion-extension). However, due to subjective determination of the starting position, the assessment of half-cycle movements (e.g, flexion only) is less reproducible.
The present study aimed to examine the reliability of measuring half-cycle cervical ROM using a spirit level for calibration.
20 healthy subjects (30 ± 12yrs, 7♂, 13♀) participated in the randomized, controlled, cross-over trial. In two testing sessions with one week of wash-out in between, cervical ROM was measured by means of an ultrasonic 3D movement analysis system using a test-retest design (baseline and 5 min post baseline). The sessions differed with reference to the mask carrying the ultrasound markers. It was removed during the 5 min break (mask off) or not (mask on). To determine the resting position, a bull's eye spirit level was used in each measurement.
With ICC values of 0.90-0.98 (mask on, p < 0.001) and 0.90 to 0.97 (mask off, p < 0.001), both examined conditions demonstrated excellent test-retest reliability for separating the cycles regarding all movement planes.
Cervical ROM during half-cycle movements can be assessed with excellent reliability using a spirit level. In contrast to subjective determination of the starting position, analyzing complete movement planes does not increase reliability. Using a defined and objective zero positioning allows the evaluation of repositioning tasks.
关节活动度(ROM)评估是临床诊断的重要组成部分。超声运动分析已被证明在分析完整幅度(例如,屈伸)时可提供可靠的结果。然而,由于起始位置的主观确定,半周期运动(例如,仅屈曲)的评估重复性较差。
本研究旨在使用水平仪校准来检查测量半周期颈椎 ROM 的可靠性。
20 名健康受试者(30±12 岁,7 名男性,13 名女性)参加了随机、对照、交叉试验。在两个测试会话之间有一周的洗脱期,使用超声 3D 运动分析系统通过测试-再测试设计(基线和基线后 5 分钟)测量颈椎 ROM。两次测试的区别在于佩戴超声标记的面罩。在 5 分钟的休息期间(取下口罩)或不取下口罩(戴上口罩)。为了确定休息位置,在每次测量中都使用了一个瞄准镜水平仪。
使用 ICC 值为 0.90-0.98(戴口罩,p<0.001)和 0.90 至 0.97(不戴口罩,p<0.001),两种检查条件在所有运动平面上都表现出极好的测试-再测试可靠性,可用于区分周期。
使用水平仪可以极好地评估半周期运动中的颈椎 ROM。与起始位置的主观确定相比,分析完整的运动平面不会增加可靠性。使用定义和客观的零定位可以评估重新定位任务。