J Refract Surg. 2017 Sep 1;33(9):617-624. doi: 10.3928/1081597X-20170620-02.
To determine the effect of lowering femtosecond laser energy on the surface quality of the intrastromal interface during small incision lenticule extraction (SMILE).
Forty age- and diopter-matched female patients (40 eyes) with moderate myopia received SMILE with different energy levels (100 to 150 nJ) and fixed spot separation (4.5 μm). Five human corneal lenticules from each energy group were evaluated by atomic force microscopy and scanning electron microscopy (SEM). Both anterior and posterior surface characteristics of the lenticules were assessed.
All measurements of surface roughness were approximately three times higher and in the anterior and posterior surface of the lenticules with the energy level of 150 nJ than with 100 nJ (P < .001). Furthermore, atomic force microscopy analysis found that energy differences of 15 nJ or more made a significant difference in surface roughness at energy levels of 115 nJ or higher. Interestingly, there was no significant difference in all roughness values of both surfaces among the 100, 105, and 110 nJ groups. In addition, all values of surface roughness were significantly positively correlated with laser energy for both anterior and posterior surfaces of the lenticule (P < .001). Consistent with atomic force microscopy results, SEM also showed that the SMILE lenticules in the higher laser energy group had more irregular surfaces.
Lowering laser energy levels can improve surface quality of the lenticule of SMILE. To achieve better visual outcomes with faster recovery after the procedure, it is recommended to reduce the laser energy to less than 115 nJ at a spot separation of 4.5 μm. [J Refract Surg. 2017;33(9):617-624.].
确定在小切口飞秒透镜切除术(SMILE)中降低飞秒激光能量对基质内界面表面质量的影响。
40例年龄和屈光度匹配的中度近视女性患者(40只眼)接受了不同能量水平(100至150 nJ)和固定光斑间距(4.5μm)的SMILE手术。通过原子力显微镜和扫描电子显微镜(SEM)对每个能量组的5个人角膜透镜进行评估。评估了透镜的前表面和后表面特征。
在透镜的前表面和后表面,150 nJ能量水平下的表面粗糙度测量值均约为100 nJ能量水平下的三倍(P <.001)。此外,原子力显微镜分析发现,在115 nJ或更高的能量水平下,15 nJ或更高的能量差异会使表面粗糙度产生显著差异。有趣的是,100、105和110 nJ组的两个表面的所有粗糙度值均无显著差异。此外,透镜前表面和后表面的所有表面粗糙度值均与激光能量呈显著正相关(P <.001)。与原子力显微镜结果一致,SEM也显示较高激光能量组的SMILE透镜表面更不规则。
降低激光能量水平可改善SMILE透镜的表面质量。为了在术后更快恢复并获得更好的视觉效果,建议在光斑间距为4.5μm时将激光能量降低至115 nJ以下。[《屈光手术杂志》。2017;33(9):617 - 624。]