Suppr超能文献

使用环境扫描电子显微镜评估 SMILE 术后人眼角膜透镜的表面质量。

Surface quality of human corneal lenticules after SMILE assessed using environmental scanning electron microscopy.

出版信息

J Refract Surg. 2014 Jun;30(6):388-93. doi: 10.3928/1081597X-20140513-01.

Abstract

PURPOSE

To assess the cut quality of the anterior and posterior surfaces of intrastromal refractive lenticules removed during small incision lenticule extraction (SMILE).

METHODS

The VisuMax femtosecond laser (500 kHz; Carl Zeiss Meditec, Dublin, CA) was used to perform SMILE on 8 eyes of 5 individuals to correct only myopia (no cylinder). The cut energy index was 26 (equivalent to an energy of 130 nJ) with a 2.5 × 2.5 μm spot/track separation. The lenticule diameter was 6.5 mm with a minimum edge thickness of 15 μm and the cap diameter was 7.3 mm with an intended thickness of 120 μm. After laser treatment, the lenticule was loosened with a spatula and removed with forceps. The extracted lenticules were placed in 2% formalin and sent for imaging with an environmental scanning electron microscope. Images of the anterior and posterior surfaces of the lenticules were obtained at multiple magnifications (100×, 250×, and 500×). Surface quality was evaluated by an investigator who specializes in electron microscopy using three criteria: overall surface regularity, percent of surface irregularity, and position of irregular area.

RESULTS

Both the anterior and posterior surfaces of the extracted lenticules were smooth and absent of surface irregularities. The cut edges also appeared uniform. Jagged edges were seen in several images, but were clearly caused by the forceps during extraction.

CONCLUSIONS

Using the VisuMax laser to perform SMILE produces smooth cuts absent of surface irregularities.

摘要

目的

评估在微切口角膜基质透镜取出术(SMILE)中切除的基质内屈光性透镜的前、后表面的切割质量。

方法

使用 VisuMax 飞秒激光(500 kHz;卡尔蔡司公司,都柏林,加利福尼亚州)对 5 名个体的 8 只眼进行 SMILE 手术,仅矫正近视(无散光)。切割能量指数为 26(相当于 130 nJ 的能量),光斑/轨道分离为 2.5×2.5 μm。透镜直径为 6.5mm,最小边缘厚度为 15μm,帽直径为 7.3mm,预期厚度为 120μm。激光治疗后,用解剖刀松解透镜,用镊子取出。提取的透镜放入 2%福尔马林溶液中,并通过环境扫描电子显微镜进行成像。在多个放大倍数(100×、250×和 500×)下获取透镜前、后表面的图像。使用三个标准:整体表面规则度、表面不规则度百分比和不规则区域位置,由专门从事电子显微镜的研究人员评估表面质量。

结果

提取的透镜的前、后表面均光滑,无表面不规则。切割边缘也显得均匀。在几张图像中可以看到锯齿状边缘,但这显然是在提取过程中镊子造成的。

结论

使用 VisuMax 激光进行 SMILE 可产生无表面不规则的光滑切口。