Suppr超能文献

脑脊液样本的自动细胞计数:GloCyte 系统的多中心性能评估。

Automated cell counts on CSF samples: A multicenter performance evaluation of the GloCyte system.

机构信息

Department of Pathology and Cell Biology, Columbia University College of Physicians and Surgeons, New York, NY, USA.

NewYork-Presbyterian Hospital, New York, NY, USA.

出版信息

Int J Lab Hematol. 2018 Feb;40(1):56-65. doi: 10.1111/ijlh.12728. Epub 2017 Sep 7.

Abstract

OBJECTIVES

Automated cell counters have replaced manual enumeration of cells in blood and most body fluids. However, due to the unreliability of automated methods at very low cell counts, most laboratories continue to perform labor-intensive manual counts on many or all cerebrospinal fluid (CSF) samples. This multicenter clinical trial investigated if the GloCyte System (Advanced Instruments, Norwood, MA), a recently FDA-approved automated cell counter, which concentrates and enumerates red blood cells (RBCs) and total nucleated cells (TNCs), is sufficiently accurate and precise at very low cell counts to replace all manual CSF counts.

METHODS

The GloCyte System concentrates CSF and stains RBCs with fluorochrome-labeled antibodies and TNCs with nucleic acid dyes. RBCs and TNCs are then counted by digital image analysis. Residual adult and pediatric CSF samples obtained for clinical analysis at five different medical centers were used for the study. Cell counts were performed by the manual hemocytometer method and with the GloCyte System following the same protocol at all sites. The limits of the blank, detection, and quantitation, as well as precision and accuracy of the GloCyte, were determined.

RESULTS

The GloCyte detected as few as 1 TNC/μL and 1 RBC/μL, and reliably counted as low as 3 TNCs/μL and 2 RBCs/μL. The total coefficient of variation was less than 20%. Comparison with cell counts obtained with a hemocytometer showed good correlation (>97%) between the GloCyte and the hemocytometer, including at very low cell counts.

CONCLUSIONS

The GloCyte instrument is a precise, accurate, and stable system to obtain red cell and nucleated cell counts in CSF samples. It allows for the automated enumeration of even very low cell numbers, which is crucial for CSF analysis. These results suggest that GloCyte is an acceptable alternative to the manual method for all CSF samples, including those with normal cell counts.

摘要

目的

自动化细胞计数器已取代血液和大多数体液中细胞的手动计数。然而,由于自动化方法在非常低的细胞计数下不可靠,大多数实验室仍继续对许多或所有脑脊液(CSF)样本进行费力的手动计数。这项多中心临床试验研究了最近获得 FDA 批准的自动化细胞计数器 GloCyte 系统(Advanced Instruments,马萨诸塞州诺伍德)是否在非常低的细胞计数下足够准确和精确,以替代所有手动 CSF 计数。

方法

GloCyte 系统浓缩 CSF,并使用荧光标记的抗体对 RBC 进行染色,使用核酸染料对 TNC 进行染色。然后通过数字图像分析对 RBC 和 TNC 进行计数。该研究使用了来自五个不同医疗中心用于临床分析的剩余成人和儿科 CSF 样本。在所有地点都遵循相同方案,通过手动血球计数器方法和 GloCyte 系统进行细胞计数。确定了 GloCyte 的空白、检测和定量极限,以及精密度和准确性。

结果

GloCyte 可检测低至 1 个 TNC/μL 和 1 个 RBC/μL 的细胞,并且可靠地计数低至 3 个 TNC/μL 和 2 个 RBC/μL 的细胞。总变异系数小于 20%。与使用血球计数器获得的细胞计数相比,GloCyte 与血球计数器之间具有良好的相关性(>97%),包括在非常低的细胞计数下。

结论

GloCyte 仪器是一种精确、准确且稳定的系统,可用于 CSF 样本中的红细胞和有核细胞计数。它允许对甚至非常低的细胞数量进行自动计数,这对 CSF 分析至关重要。这些结果表明,GloCyte 是一种可接受的替代手动方法,适用于所有 CSF 样本,包括细胞计数正常的样本。