J Diabetes Sci Technol. 2017 Nov;11(6):1163-1164. doi: 10.1177/1932296817730084. Epub 2017 Sep 14.

Abstract

The article titled "Seven Year Surveillance of the Clinical Performance of a Blood Glucose Test-Strip Product" by Setford and coworkers in this issue of Journal of Diabetes Science and Technology is an impressive study showing that over 7 years in three clinics, using multiple reagent lots, a total of 73 600 samples met the ISO 15197 2015 standard with no results in the D or E zones of a Parkes glucose meter error grid. Three requirements are suggested for a clinically acceptable glucose meter. The authors provide strong evidence for meeting two requirements but fail to provide summarized data about the number of nonnumeric results. Finally, the authors overstate some results, called "spin" by some which is not necessary. The superb results should stand on their own.

摘要

Setford及其同事在本期《糖尿病科学与技术杂志》上发表的题为《血糖试纸产品临床性能的七年监测》的文章是一项令人印象深刻的研究,表明在三家诊所进行的7年多时间里,使用多个试剂批次,共有73600个样本符合ISO 15197 2015标准,在帕克斯血糖仪误差网格的D区或E区没有出现结果。文章提出了临床可接受血糖仪的三个要求。作者为满足其中两个要求提供了有力证据,但未能提供关于非数字结果数量的汇总数据。最后,作者夸大了一些结果,有些人称之为“夸大”,这是没有必要的。卓越的结果应该靠自身站住脚。

相似文献

1
Analysis of "Seven Year Surveillance of the Clinical Performance of a Blood Glucose Test-Strip Product".
J Diabetes Sci Technol. 2017 Nov;11(6):1163-1164. doi: 10.1177/1932296817730084. Epub 2017 Sep 14.
2
Why the Diabetes Technology Society Surveillance Protocol for Glucose Meters Needs to Be Revised.
J Diabetes Sci Technol. 2017 Nov;11(6):1247-1249. doi: 10.1177/1932296817702926. Epub 2017 Apr 5.
3
Performance of two updated blood glucose monitoring systems: an evaluation following ISO 15197:2013.
Curr Med Res Opin. 2016 May;32(5):847-55. doi: 10.1185/03007995.2016.1146666. Epub 2016 Feb 23.
4
The surveillance error grid.
J Diabetes Sci Technol. 2014 Jul;8(4):658-72. doi: 10.1177/1932296814539589. Epub 2014 Jun 13.
5
Seven-Year Surveillance of the Clinical Performance of a Blood Glucose Test Strip Product.
J Diabetes Sci Technol. 2017 Nov;11(6):1155-1162. doi: 10.1177/1932296817703133. Epub 2017 Apr 13.
6
Introduction of a Novel Smartphone-Coupled Blood Glucose Monitoring System.
J Diabetes Sci Technol. 2017 Nov;11(6):1231-1233. doi: 10.1177/1932296817706594. Epub 2017 May 1.
7
Accuracy Evaluation of a CE-Marked Glucometer System for Self-Monitoring of Blood Glucose With Three Reagent Lots Following ISO 15197:2013.
J Diabetes Sci Technol. 2017 May;11(3):635-636. doi: 10.1177/1932296816670401. Epub 2016 Oct 4.
8
The clinical performance and ease of use of a blood glucose meter that uses a 10-test disk.
Diabetes Technol Ther. 2004 Aug;6(4):495-502. doi: 10.1089/1520915041705983.

本文引用的文献

1
Overinterpretation of Research Findings: Evidence of "Spin" in Systematic Reviews of Diagnostic Accuracy Studies.
Clin Chem. 2017 Aug;63(8):1353-1362. doi: 10.1373/clinchem.2017.271544. Epub 2017 Jun 12.
2
Seven-Year Surveillance of the Clinical Performance of a Blood Glucose Test Strip Product.
J Diabetes Sci Technol. 2017 Nov;11(6):1155-1162. doi: 10.1177/1932296817703133. Epub 2017 Apr 13.
3
Performance of Cleared Blood Glucose Monitors.
J Diabetes Sci Technol. 2015 Jul;9(4):895-910. doi: 10.1177/1932296815584797. Epub 2015 Jun 8.
4
The surveillance error grid.
J Diabetes Sci Technol. 2014 Jul;8(4):658-72. doi: 10.1177/1932296814539589. Epub 2014 Jun 13.
5
The chronic injury glucose error grid: a tool to reduce diabetes complications.
J Diabetes Sci Technol. 2015 Jan;9(1):149-52. doi: 10.1177/1932296814554415. Epub 2014 Oct 14.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验