Suppr超能文献

从仰卧位到站立位时髋臼杯方向的变化及其对超高交联聚乙烯磨损的影响。

Change in Acetabular Cup Orientation From Supine to Standing Position and Its Effect on Wear of Highly Crosslinked Polyethylene.

机构信息

Department of Surgery, Schulich School of Medicine & Dentistry, Western University and London Health Sciences Centre, London, Ontario, Canada; Surgical Innovation Program, Lawson Health Research Institute, London, Ontario, Canada; Imaging Research Laboratories, Robarts Research Institute, London, Ontario, Canada.

Department of Surgery, Schulich School of Medicine & Dentistry, Western University and London Health Sciences Centre, London, Ontario, Canada.

出版信息

J Arthroplasty. 2018 Jan;33(1):263-267. doi: 10.1016/j.arth.2017.08.016. Epub 2017 Aug 31.

Abstract

BACKGROUND

The purpose of this study is to measure acetabular cup position and wear of the highly crosslinked polyethylene liner in the supine and standing position for patients at a minimum of 10 years after the operation.

METHODS

A total of 38 patients were recruited at a mean of 12.5 years after the operation. All patients received a single acetabular cup design with a highly crosslinked liner and a 28-mm cobalt-chromium femoral head. Patients underwent supine and standing radiostereometric examinations in which the X-ray sources and detectors were positioned to obtain an anterior-posterior and cross-table lateral radiograph. Acetabular cup position and the three-dimensional wear rate were measured from the radiographs, and outcome scores were recorded for each patient.

RESULTS

Anteversion significantly increased (P < .0001) a mean of 12° from supine (15.1° ± 10.4°) to standing (27.2° ± 10.5°) position. Inclination also significantly increased (P = .001) a mean of 2° from supine (44.4° ± 6.8°) to standing (46.3° ± 7.7°) position. There was no difference (P = .093) in wear rate between supine (0.067 ± 0.070 mm/y) and standing (0.073 ± 0.074 mm/y) positions. There were no correlations between cup orientation and wear rate in either position.

CONCLUSION

Highly crosslinked polyethylene is a forgiving bearing material. Although adherence to the traditional acetabular position target zone is recommended, ensuring hip stability and consideration of the patient's functional position are also important objectives to consider for the acetabular position.

摘要

背景

本研究的目的是测量术后至少 10 年患者在仰卧位和站立位时髋臼杯位置和高交联聚乙烯衬垫的磨损情况。

方法

共招募了 38 名患者,平均手术时间为 12.5 年。所有患者均接受了单髋臼杯设计,使用高交联衬垫和 28 毫米钴铬股骨头。患者接受仰卧位和站立位放射立体测量检查,其中 X 射线源和探测器的位置设置为获得前后位和交叉位侧位 X 线片。从 X 线片中测量髋臼杯位置和三维磨损率,并记录每位患者的结果评分。

结果

前倾角从仰卧位(15.1°±10.4°)显著增加(P<0.0001)至站立位(27.2°±10.5°),平均增加 12°。倾斜度也从仰卧位(44.4°±6.8°)显著增加(P=0.001)至站立位(46.3°±7.7°),平均增加 2°。仰卧位(0.067±0.070mm/y)和站立位(0.073±0.074mm/y)之间的磨损率无差异(P=0.093)。在任何位置,杯状取向与磨损率之间均无相关性。

结论

高交联聚乙烯是一种宽容的承载材料。尽管建议遵循传统的髋臼位置目标区域,但确保髋关节稳定性和考虑患者的功能位置也是髋臼位置的重要目标。