Camison L, Bykowski M, Lee W W, Carlson J C, Roosenboom J, Goldstein J A, Losee J E, Weinberg S M
Division of Pediatric Plastic Surgery, Children's Hospital of Pittsburgh, Pittsburgh, Pennsylvania, USA.
Department of Biostatistics, University of Pittsburgh, Pittsburgh, Pennsylvania, USA.
Int J Oral Maxillofac Surg. 2018 Mar;47(3):403-410. doi: 10.1016/j.ijom.2017.08.008. Epub 2017 Sep 14.
Three-dimensional (3D) surface imaging using stereophotogrammetry has become increasingly popular in clinical settings, offering advantages for surgical planning and outcome evaluation. The handheld Vectra H1 is a low-cost, highly portable system that offers several advantages over larger stationary cameras, but independent technical validation is currently lacking. In this study, 3D facial images of 26 adult participants were captured with the Vectra H1 system and the previously validated 3dMDface system. Using error magnitude statistics, 136 linear distances were compared between cameras. In addition, 3D facial surfaces from each system were registered, heat maps generated, and global root mean square (RMS) error calculated. The 136 distances were highly comparable across the two cameras, with an average technical error of measurement (TEM) value of 0.84mm (range 0.19-1.54mm). The average RMS value of the 26 surface-to-surface comparisons was 0.43mm (range 0.33-0.59mm). In each case, the vast majority of the facial surface differences were within a ±1mm threshold. Areas exceeding ±1mm were generally limited to facial regions containing hair or subject to facial microexpressions. These results indicate that 3D facial surface images acquired with the Vectra H1 system are sufficiently accurate for most clinical applications.
使用立体摄影测量法的三维(3D)表面成像在临床环境中越来越受欢迎,为手术规划和结果评估提供了优势。手持式Vectra H1是一种低成本、高度便携的系统,与大型固定相机相比具有多个优势,但目前缺乏独立的技术验证。在本研究中,使用Vectra H1系统和先前经过验证的3dMDface系统采集了26名成年参与者的3D面部图像。使用误差幅度统计,比较了两台相机之间的136个线性距离。此外,对每个系统的3D面部表面进行配准,生成热图,并计算全局均方根(RMS)误差。两台相机的136个距离具有高度可比性,平均测量技术误差(TEM)值为0.84mm(范围为0.19 - 1.54mm)。26次表面到表面比较的平均RMS值为0.43mm(范围为0.33 - 0.59mm)。在每种情况下,绝大多数面部表面差异都在±1mm阈值内。超过±1mm的区域通常限于含有毛发或存在面部微表情的面部区域。这些结果表明,用Vectra H1系统采集的3D面部表面图像对于大多数临床应用来说足够准确。