用于近距离放射治疗中源追踪的时间分辨体内剂量测定法。

Time-resolved in vivo dosimetry for source tracking in brachytherapy.

作者信息

Johansen Jacob Graversen, Rylander Susanne, Buus Simon, Bentzen Lise, Hokland Steffen Bjerre, Søndergaard Christian Skou, With Anders Karl Mikael, Kertzscher Gustavo, Tanderup Kari

机构信息

Department of Oncology, Aarhus University Hospital, Aarhus C, Denmark.

Department of Oncology, Aarhus University Hospital, Aarhus C, Denmark.

出版信息

Brachytherapy. 2018 Jan-Feb;17(1):122-132. doi: 10.1016/j.brachy.2017.08.009. Epub 2017 Sep 22.

Abstract

PURPOSE

The purpose of this article is to demonstrate that brachytherapy source tracking can be realized with in vivo dosimetry. This concept could enable real-time treatment monitoring.

METHODS

In vivo dosimetry was incorporated in the clinical routine during high-dose-rate prostate brachytherapy at Aarhus University Hospital. The dosimetry was performed with a radioluminescent crystal positioned in a dedicated brachytherapy needle in the prostate. The dose rate was recorded every 50-100 ms during treatment and analyzed retrospectively. The measured total delivered dose and dose rates for each dwell position with dwell times >0.7 s were compared with expected values. Furthermore, the distance between the source and dosimeter, which was derived from the measured dose rates, was compared with expected values. The measured dose rate pattern in each needle was used to determine the most likely position of the needle relative to the dosimeter.

RESULTS

In total, 305 needles and 3239 dwell positions were analyzed based on 20 treatments. The measured total doses differed from the expected values by -4.7 ± 8.4% (1SD) with range (-17% to 12%). It was possible to determine needle shifts for 304 out of 305 needles. The mean radial needle shift between imaging and treatment was 0.2 ± 1.1 mm (1SD), and the mean longitudinal shift was 0.3 ± 2.0 mm (1SD).

CONCLUSION

Time-resolved in vivo dosimetry can be used to provide geometric information about the treatment progression of afterloading brachytherapy. This information may provide a clear indication of errors and uncertainties during a treatment and, therefore, enables real-time treatment monitoring.

摘要

目的

本文旨在证明近距离放射治疗源跟踪可通过体内剂量测定来实现。这一概念能够实现实时治疗监测。

方法

在奥胡斯大学医院进行高剂量率前列腺近距离放射治疗期间,将体内剂量测定纳入临床常规操作。剂量测定通过置于前列腺专用近距离放射治疗针内的放射性发光晶体进行。在治疗过程中每50 - 100毫秒记录一次剂量率,并进行回顾性分析。将每个驻留时间>0.7秒的驻留位置的测量总输送剂量和剂量率与预期值进行比较。此外,将根据测量剂量率得出的源与剂量计之间的距离与预期值进行比较。利用每根针中测量的剂量率模式来确定针相对于剂量计最可能的位置。

结果

基于20次治疗,共分析了305根针和3239个驻留位置。测量的总剂量与预期值相差 -4.7 ± 8.4%(1标准差),范围为(-17%至12%)。305根针中有304根能够确定针的移位。成像与治疗之间针的平均径向移位为0.2 ± 1.1毫米(1标准差),平均纵向移位为0.3 ± 2.0毫米(1标准差)。

结论

时间分辨体内剂量测定可用于提供关于后装近距离放射治疗治疗进程的几何信息。该信息可明确指示治疗过程中的误差和不确定性,从而实现实时治疗监测。