评估脉冲氙气紫外线消毒系统以减少手术室中的细菌污染。

Evaluation of a pulsed xenon ultraviolet disinfection system to decrease bacterial contamination in operating rooms.

作者信息

El Haddad Lynn, Ghantoji Shashank S, Stibich Mark, Fleming Jason B, Segal Cindy, Ware Kathy M, Chemaly Roy F

机构信息

Department of Infectious Diseases, Infection Control and Employee Health, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Boulevard, Unit 1460, Houston, TX, 77030-4095, USA.

Xenex Disinfection Services, San Antonio, TX, USA.

出版信息

BMC Infect Dis. 2017 Oct 10;17(1):672. doi: 10.1186/s12879-017-2792-z.

Abstract

BACKGROUND

Environmental cleanliness is one of the contributing factors for surgical site infections in the operating rooms (ORs). To decrease environmental contamination, pulsed xenon ultraviolet (PX-UV), an easy and safe no-touch disinfection system, is employed in several hospital environments. The positive effect of this technology on environmental decontamination has been observed in patient rooms and ORs during the end-of-day cleaning but so far, no study explored its feasibility between surgical cases in the OR.

METHODS

In this study, 5 high-touch surfaces in 30 ORs were sampled after manual cleaning and after PX-UV intervention mimicking between-case cleaning to avoid the disruption of the ORs' normal flow. The efficacy of a 1-min, 2-min, and 8-min cycle were tested by measuring the surfaces' contaminants by quantitative cultures using Tryptic Soy Agar contact plates.

RESULTS

We showed that combining standard between-case manual cleaning of surfaces with a 2-min cycle of disinfection using a portable xenon pulsed ultraviolet light germicidal device eliminated at least 70% more bacterial load after manual cleaning.

CONCLUSIONS

This study showed the proof of efficacy of a 2-min cycle of PX-UV in ORs in eliminating bacterial contaminants. This method will allow a short time for room turnover and a potential reduction of pathogen transmission to patients and possibly surgical site infections.

摘要

背景

环境清洁是手术室手术部位感染的促成因素之一。为减少环境污染,脉冲氙气紫外线(PX-UV)这种简便安全的非接触式消毒系统已在多个医院环境中使用。在病房和手术室的终末清洁过程中已观察到该技术对环境净化的积极作用,但迄今为止,尚无研究探讨其在手术室手术病例之间使用的可行性。

方法

在本研究中,对30间手术室中的5个高频接触表面在人工清洁后以及模拟病例间清洁的PX-UV干预后进行采样,以避免干扰手术室的正常流程。通过使用胰蛋白胨大豆琼脂接触平板进行定量培养来测量表面污染物,测试1分钟、2分钟和8分钟周期的消毒效果。

结果

我们发现,将标准的病例间表面人工清洁与使用便携式氙气脉冲紫外线杀菌装置进行2分钟周期的消毒相结合,在人工清洁后可消除至少70%以上的细菌负荷。

结论

本研究证明了在手术室中使用2分钟周期的PX-UV消除细菌污染物的有效性。这种方法将缩短房间周转时间,并有可能减少病原体传播给患者以及可能减少手术部位感染。