Amare Meareg, Menkir Getahun
Bahir Dar University, Department of Chemistry, Bahir Dar, Ethiopia.
Enjibara Teachers College, Enjibara, Amhara Regional State, Ethiopia.
Heliyon. 2017 Oct 3;3(10):e00417. doi: 10.1016/j.heliyon.2017.e00417. eCollection 2017 Oct.
A new method for determination of salbutamol sulfate has been developed using poly(4-amino-3-hydroxynaphthalene sulfonic acid/GCE. Cyclic voltammetric investigation of the electrochemical behavior of salbutamol sulfate at the polymer modified glassy carbon unveiled electrocatalytic activity of the modifier towards irreversible oxidation of salbutamol sulfate. Dependence of peak current predominantly on scan rate than on square root of scan rate, and peak potential shift with pH demonstrated that oxidation of salbutamol sulfate at the polymer modified electrode follows adsorption reaction kinetics with proton participation. Under optimized solution and differential pulse voltammetric parameters, the oxidative peak current showed linear dependence on salbutamol sulfate concentration in the range 0.2 to 8 μM with method detection limit (3s/m) and determination coefficient (R) of 6.8 × 10 M and 0.99786, respectively. Low method detection limit, relatively wide linear range, and recovery results of spiked standard salbutamol sulfate in syrup samples in the range 96.7-98.9% validated the method for determination of salbutamol sulfate in pharmaceutical formulations. Differential pulse voltammetric analysis of salbutamol sulfate syrup formulation for its salbutamol sulfate content revealed 98.8 to 99.3% of the labeled value confirming the applicability of the developed method for determination of salbutamol sulfate in real samples.
已开发出一种使用聚(4-氨基-3-羟基萘磺酸)修饰玻碳电极测定硫酸沙丁胺醇的新方法。对硫酸沙丁胺醇在聚合物修饰玻碳电极上的电化学行为进行循环伏安研究,揭示了修饰剂对硫酸沙丁胺醇不可逆氧化的电催化活性。峰电流主要依赖于扫描速率而非扫描速率的平方根,且峰电位随pH值变化,这表明在聚合物修饰电极上硫酸沙丁胺醇的氧化遵循有质子参与的吸附反应动力学。在优化的溶液和差分脉冲伏安参数条件下,氧化峰电流在0.2至8 μM范围内对硫酸沙丁胺醇浓度呈线性依赖,方法检测限(3s/m)和测定系数(R)分别为6.8×10⁻⁸ M和0.99786。低方法检测限、相对较宽的线性范围以及糖浆样品中加标硫酸沙丁胺醇标准品的回收率在96.7 - 98.9%范围内,验证了该方法用于药物制剂中硫酸沙丁胺醇测定的可行性。对硫酸沙丁胺醇糖浆制剂中硫酸沙丁胺醇含量进行差分脉冲伏安分析,结果显示为标示值的98.8%至99.3%,证实了所开发方法用于实际样品中硫酸沙丁胺醇测定的适用性。