Gahrn-Hansen B, Alstrup P, Dessau R, Fuursted K, Knudsen A, Olsen H, Oxhøj H, Petersen A R, Siboni A, Siboni K
Department of Clinical Microbiology, Odense University Hospital, Denmark.
J Hosp Infect. 1988 Jul;12(1):1-6. doi: 10.1016/0195-6701(88)90115-6.
Achromobacter xylosoxidans contaminating transducers caused 15 cases of hospital infection. In the eight patients with bacteraemia the interval from inoculation to fever was an average of 6.6 days. All the infected patients recovered. Computerization of laboratory records allowed retrieval of previous isolates, and review of clinical records focused the problem on patients with cardiac and aortic diseases. The problem arose from the re-use of disposable equipment after disinfection with a benzalcone.
木糖氧化无色杆菌污染传感器导致了15例医院感染。在8例菌血症患者中,从接种到发热的间隔平均为6.6天。所有感染患者均康复。实验室记录的计算机化使得能够检索先前的分离株,而临床记录的回顾将问题聚焦于患有心脏和主动脉疾病的患者。该问题是由于使用苯扎氯铵消毒后重复使用一次性设备引起的。