手持式肌动计评估急性脑卒中患者肌肉张力的日内组内信度。

Between-days intra-rater reliability with a hand held myotonometer to quantify muscle tone in the acute stroke population.

机构信息

Department of Rehabilitation Medicine, Guangdong Engineering and Technology Research Center for Rehabilitation Medicine and Translation, The First Affiliated Hospital, Sun Yat-sen University, Guangzhou, 510080, China.

Department of Electronic and Electrical Engineering, University College London, Torrington Place, London, WC1E 7JE, England.

出版信息

Sci Rep. 2017 Oct 26;7(1):14173. doi: 10.1038/s41598-017-14107-3.

Abstract

A myotonometer can objectively quantify changes in muscle tone. The between-days intra-rater reliability in a ward setting for the acute stroke population remains unknown. This study aimed to investigate the device's between-days intra-rater reliability when used in a ward setting for acute stroke participants. Muscle tone of biceps brachii, brachioradialis, rectus femoris, and tibialis anterior was recorded in the ward at bedside by one physiotherapist on two consecutive days. This study included participants who were within 1 month of their first stroke occurrence. Participants who were medically unstable or who suffered from brain stem injury were excluded. Reliability was assessed by the intraclass correlation coefficient (ICC), standard error of measurement (SEM), smallest real difference (SRD), and the Bland-Altman limits of agreement. The results indicated excellent between-days intra-rater reliability (ICC > 0.75). SEM and SRD show small differences between measurements. The Bland-Altman analysis indicated a tendency of overestimation of the rectus femoris. MyotonPRO demonstrated acceptable reliability when used in a ward setting in those patients with acute stroke. However, results should be interpreted with caution, due to the limitations of the study and the varying level of consistency observed between different muscles.

摘要

肌振器可以客观地量化肌肉张力的变化。在急性中风患者的病房环境中,该设备的两天内组内可靠性尚不清楚。本研究旨在调查该设备在急性中风患者的病房环境中的两天内组内可靠性。由一名物理治疗师在两天内在床边记录肱二头肌、肱桡肌、股四头肌和胫骨前肌的肌肉张力。本研究纳入了首次中风发作后 1 个月内的患者。排除了病情不稳定或患有脑干损伤的患者。可靠性通过组内相关系数(ICC)、测量误差(SEM)、最小实际差异(SRD)和 Bland-Altman 一致性界限进行评估。结果表明,该设备具有极好的两天内组内可靠性(ICC>0.75)。SEM 和 SRD 显示出测量值之间的微小差异。Bland-Altman 分析表明股四头肌存在高估的趋势。肌振器在急性中风患者的病房环境中具有可接受的可靠性。然而,由于研究的局限性以及不同肌肉之间观察到的一致性程度不同,结果应谨慎解释。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/2dc2/5658427/2e5c7cd502fe/41598_2017_14107_Fig1_HTML.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验