Lassaletta Luis, Polak Marek, Huesers Jan, Díaz-Gómez Miguel, Calvino Miryam, Varela-Nieto Isabel, Gavilán Javier
*Department of Otolaryngology, "La Paz" University Hospital †Centro de Investigación Biomédica en Red de Enfermedades Raras (CIBERER-U761) ‡Instituto de Investigación Hospital Universitario La Paz (IdiPAZ), Madrid, Spain §Med-El Corporation, Insbruck, Austria ||Instituto de Investigaciones Biomédicas "Alberto Sols" CSIC-UAM, Madrid, Spain.
Otol Neurotol. 2017 Dec;38(10):e413-e420. doi: 10.1097/MAO.0000000000001584.
To use an intracochlear test electrode to assess the integrity and the functionality of the auditory nerve in cochlear implant (CI) recipients and to compare electrical auditory brainstem responses (eABR) via the test electrode with the eABR responses with the CI.
Otolaryngology department, tertiary referral hospital.
Ten subjects (age at implantation 55 yr, range, 19-72) were subsequently implanted with a MED-EL CONCERTO CI on the side without any useful residual hearing.
Following identification of the round window (RW), the test electrode was inserted in the cochlea previous to cochlear implantation.
To assess the quality of an eABR waveform, scoring criteria from Walton et al. (2008) were chosen. The waveforms in each session were classified by detecting waves III and V by the algorithm and visual assessment of the waveform. Speech performance was evaluated with monosyllables, disyllables, and sentence recognition tests.
It was possible to evoke electrical stimulation responses along with both the test electrode and the CI in all subjects. No significant differences in latencies or amplitudes after stimulation were found between the test electrode and the CI. All subjects obtained useful hearing with their CI and use their implants daily.
The intracochlear test electrode may be suitable to test the integrity of the auditory nerve by recording eABR signals. This allows for further research on the status of the auditory nerve after tumor removal and correlation with auditory performance.
使用耳蜗内测试电极评估人工耳蜗(CI)植入者听神经的完整性和功能,并比较通过测试电极获得的电诱发听性脑干反应(eABR)与通过CI获得的eABR反应。
三级转诊医院耳鼻喉科。
10名受试者(植入时年龄55岁,范围19 - 72岁)随后在无任何有用残余听力的一侧植入了MED-EL CONCERTO CI。
在识别出圆窗(RW)后,在植入人工耳蜗之前将测试电极插入耳蜗。
为评估eABR波形质量,选用了Walton等人(2008年)的评分标准。通过算法检测波III和波V并对波形进行视觉评估,对每个阶段的波形进行分类。通过单音节、双音节和句子识别测试评估言语表现。
在所有受试者中,使用测试电极和CI均能诱发电刺激反应。测试电极和CI之间在刺激后潜伏期或波幅方面未发现显著差异。所有受试者通过其CI获得了有用听力并每天使用植入物。
耳蜗内测试电极可能适用于通过记录eABR信号来测试听神经的完整性。这有助于对肿瘤切除后听神经的状态进行进一步研究,并与听觉表现进行相关性分析。