Biomed Opt Express. 2017 Sep 21;8(10):4621-4628. doi: 10.1364/BOE.8.004621. eCollection 2017 Oct 1.
To study the optical changes on hydrogel-silicone intraocular lenses (IOLs) resulting from loading them with dexamethasone. We used prototype hydrogel(pHEMA)-silicone IOLs and loaded the matrices with an anti-inflammatory drug (dexamethasone). The optical properties we analyzed experimentally were a) modulation transfer function (MTF); b) spectral transmission; c) diopter power. These determinations were performed on drug-loaded IOLs, IOLs that had released the drug, and IOLs that had not been drug-loaded. Loading a hydrogel-silicone IOL with dexamethasone results in impairment of its optical qualities, in particular its MTF and spectral transmission, but not dioptric power. However, once the drug has been released, it almost recovers its initial optical properties.