Increase in Benign Squeaking Rate at Five-Year Follow-Up: Results of a Large Diameter Ceramic-on-Ceramic Bearing in Total Hip Arthroplasty.

机构信息

Specialist Orthopaedic Group, The Mater Clinic, North Sydney, North South Wales, Australia.

出版信息

J Arthroplasty. 2018 Apr;33(4):1210-1214. doi: 10.1016/j.arth.2017.11.044. Epub 2017 Dec 2.

Abstract

BACKGROUND

Squeaking is an on-going complication with a variable incidence of 0.5%-20.7%. The mechanism of squeaking has not been understood completely and is most likely multifactorial in nature. Previously, we have reported on a squeaking rate of 7.3% at 2 years. Our current results show a substantial increase in the squeaking incidence from 7.3% to 17.4% with the DeltaMotion articulation at the 5-year follow-up.

METHODS

Two hundred six total hip arthroplasties with a large ceramic-on-ceramic bearing were performed on 195 patients by 2 senior authors. The minimum duration of follow-up was 5 years (range 5-6.2).

RESULTS

The mean Harris Hip Score remained stable at 91.7 at 5 years compared to 92 after 2 years. However, we observed an increase in the squeaking rate from 7.3% after 2 years to 17.4% after 5 years. As in previous studies, we did not find significant differences between the silent and squeaking group in relation to age, height, weight, and femoral head diameter. In contrast, others reported no significant differences regarding range of motion and gender between the silent and the squeaking hip group at 2 years of follow-up, whereas at 5 years, the squeaking group showed a significantly higher combined range of motion and a higher relative risk of squeaking in women compared to the silent group.

CONCLUSION

While large diameter ceramic bearings may produce squeaking, especially in female patients, our mid-term clinical results of a large ceramic-on-ceramic bearing in total hip arthroplasty are encouraging. Nevertheless, long-term follow-up is recommended.

摘要

背景

squeaking 是一种持续存在的并发症,其发生率为 0.5%-20.7%,变化不定。 squeaking 的发生机制尚未完全阐明,很可能是多因素的。此前,我们报道了 2 年时 squeaking 的发生率为 7.3%。我们目前的结果显示,在 5 年随访时,DeltaMotion 关节的 squeaking 发生率从 7.3%显著增加至 17.4%。

方法

由两位资深作者对 195 名患者的 206 例全髋关节置换术使用了大陶瓷-陶瓷轴承。随访时间至少为 5 年(范围为 5-6.2 年)。

结果

与术后 2 年的 92 分相比,5 年时的平均 Harris 髋关节评分保持稳定,为 91.7 分。然而,我们观察到 squeaking 的发生率从术后 2 年的 7.3%增加到术后 5 年的 17.4%。与之前的研究一样,我们没有发现 silent 和 squeaking 组在年龄、身高、体重和股骨头直径方面存在显著差异。相比之下,其他人在 2 年随访时报告说,在 silent 和 squeaking 髋关节组之间,运动范围和性别没有显著差异,而在 5 年时,squeaking 组的联合运动范围明显更高,且女性 squeaking 的相对风险高于 silent 组。

结论

虽然大直径陶瓷轴承可能会产生 squeaking,尤其是在女性患者中,但我们对大陶瓷-陶瓷轴承在全髋关节置换术中的中期临床结果感到鼓舞。然而,建议进行长期随访。