Payne Catherine, Watt Peter, Cercignani Mara, Webborn Nick
Centre for Sport and Exercise Science and Medicine (SESAME), University of Brighton, Carlisle Road, Eastbourne, BN20 7SN, UK.
Clinical Imaging Sciences Centre, Brighton and Sussex Medical School, Falmer, BN1 9RR, UK.
Skeletal Radiol. 2018 Jun;47(6):779-784. doi: 10.1007/s00256-017-2846-8. Epub 2017 Dec 19.
To assess the reproducibility of shear wave elastography (SWE) measures in the Achilles tendon (AT) in vivo.
Shear wave velocity (SWV) of 14 healthy volunteers [7 males, 7 females; mean age 26.5 ± 3.8 years, mean height 171.6 ± 10.9 cm, mean Victorian Institute of Sports Assessment Achilles questionnaire (VISA-A) score 99.4 ± 1.2] was measured with the foot relaxed and fixed at 90°. Data were collected over five consecutive measures and 5 consecutive days.
Mean SWV values ranged from 7.91 m/s-9.56 m/s ± 0.27-0.50 m/s. Coefficient of variation (CV), correlations and intra-class correlation coefficient (ICC) scores ranged from 2.9%-6.3%, 0.4-0.7 and 0.54-0.85 respectively. No significant differences were noted for longitudinal or transverse data with respect to protocol or time and no significant differences were noted for foot position in transverse data. Significant differences in SWV values were noted between foot positions for longitudinal scanning (p = <0.05), with a relaxed foot position providing SWV values on average 0.47 m/s faster than a fixed position. Increased reproducibility was obtained with the foot relaxed. ICC between operators was 0.70 for transverse and 0.80 for longitudinal scanning.
Reproducible SWE measures were obtained over a 1-h period as well as a period of 5 consecutive days with more reliable measures obtained from a longitudinal plane using a relaxed foot position. SWE also has a high level of agreement between operators making SWE a reproducible technique for quantitatively assessing the mechanical properties of the human AT in vivo.
评估体内跟腱剪切波弹性成像(SWE)测量的可重复性。
对14名健康志愿者[7名男性,7名女性;平均年龄26.5±3.8岁,平均身高171.6±10.9厘米,平均维多利亚运动评估跟腱问卷(VISA - A)评分99.4±1.2]在足部放松并固定于90°时测量剪切波速度(SWV)。数据在连续5次测量以及连续5天内收集。
平均SWV值范围为7.91米/秒至9.56米/秒±0.27至0.50米/秒。变异系数(CV)、相关性和组内相关系数(ICC)评分分别为2.9%至6.3%、0.4至0.7和0.54至0.85。纵向或横向数据在方案或时间方面未观察到显著差异,横向数据中足部位置也未观察到显著差异。纵向扫描时,不同足部位置的SWV值存在显著差异(p = <0.05),足部放松位置的SWV值平均比固定位置快0.47米/秒。足部放松时可获得更高的可重复性。横向扫描时操作者间的ICC为0.70,纵向扫描时为0.80。
在1小时期间以及连续5天内均可获得可重复的SWE测量结果,使用足部放松位置从纵向平面获得的测量结果更可靠。SWE在操作者之间也具有高度一致性,这使得SWE成为一种可重复的技术,用于定量评估体内人跟腱的力学性能。