Payne Catherine, Webborn Nick, Watt Peter, Cercignani Mara
Centre for Sport and Exercise Science and Medicine (SESAME), University of Brighton, Carlisle Road, Eastbourne, BN20 7SN, UK.
Clinical Imaging Sciences Centre, Brighton and Sussex Medical School, Falmer, BN1 9RR, UK.
Skeletal Radiol. 2017 Jul;46(7):889-895. doi: 10.1007/s00256-017-2629-2. Epub 2017 Apr 4.
To determine the reproducibility of compression elastography (CE) when measuring strain data, a measure of stiffness of the human Achilles tendon in vivo, over consecutive measures, consecutive days and when using different foot positions.
Eight participants (4 males, 4 females; mean age 25.5 ± 2.51 years, range 21-30 years; height 173.6 ± 11.7 cm, range 156-189 cm) had five consecutive CE measurements taken on one day and a further five CE measures taken, one per day, at the same time of day, every day for a consecutive 5-day period. These 80 measurements were used to assess both the repeatability and reproducibility of the technique. Means, standard deviations, coefficient of variation (CV), Pearson correlation analysis (R) and intra-class correlation coefficients (ICC) were calculated.
For CE data, all CVs were above 53%, R values indicated no-to-weak correlations between measures at best (range 0.01-0.25), and ICC values were all classified in the poor category (range 0.00-0.11). CVs for length and diameter measures were acceptably low indicating a high level of reliability.
Given the wide variation obtained in the CE results, it was concluded that CE using this specific system has a low level of reproducibility for measuring the stiffness of the human Achilles tendon in vivo over consecutive days, consecutive measures and in different foot positions.
确定在连续测量、连续数天以及使用不同足部位置时,压缩弹性成像(CE)测量应变数据(一种体内人体跟腱刚度的测量指标)的可重复性。
八名参与者(4名男性,4名女性;平均年龄25.5±2.51岁,范围21 - 30岁;身高173.6±11.7厘米,范围156 - 189厘米)在一天内连续进行了五次CE测量,并在接下来连续5天的每天同一时间又进行了五次CE测量,每次测量一次。这80次测量用于评估该技术的重复性和可重复性。计算了均值、标准差、变异系数(CV)、Pearson相关分析(R)和组内相关系数(ICC)。
对于CE数据,所有CV均高于53%,R值表明测量之间充其量存在无至弱的相关性(范围0.01 - 0.25),并且ICC值均被归类为差(范围0.00 - 0.11)。长度和直径测量的CV低至可接受水平,表明可靠性较高。
鉴于CE结果存在广泛差异,得出结论:使用该特定系统的CE在连续数天、连续测量以及不同足部位置下测量人体跟腱在体内的刚度时,可重复性较低。