J Endod. 2018 Mar;44(3):489-495. doi: 10.1016/j.joen.2017.09.008. Epub 2017 Dec 19.
This study evaluated the shaping ability of the XP-endo Shaper (FKG Dentaire SA, La Chaux-de-Fonds, Switzerland), iRaCe (FKG Dentaire SA), and EdgeFile (EdgeEndo, Albuquerque, NM) systems using micro-computed tomographic (micro-CT) technology.
Thirty long oval-shaped canals from mandibular incisors were matched anatomically using micro-CT scanning (SkyScan1174v2; Bruker-microCT, Kontich, Belgium) and distributed into 3 groups (n = 10) according to the canal preparation protocol (ie, XP-endo Shaper, iRaCe, and EdgeFile systems). Coregistered images, before and after preparation, were evaluated for morphometric measurements of the volume, surface area, structure model index (SMI), untouched walls, area, perimeter, roundness, and diameter. Data were statistically compared between groups using the 1-way analysis of variance post hoc Tukey test and within groups with the paired sample t test (α = 5%).
Within groups, preparation significantly increased all tested parameters (P < .05). No statistical difference was observed in the mean percentage increase of the volume (〜52%) and surface area (10.8%-14.2%) or the mean percentage of the remaining unprepared canal walls between groups (8.17%-9.83%) (P > .05). The XP-endo Shaper significantly altered the overall geometry of the root canal to a more conical shape (SMI = 2.59) when compared with the other groups (P < .05). After preparation protocols, changes in area, perimeter, roundness, and minor and major diameters of the root canals in the 5 mm of the root apex showed no difference between groups (P > .05).
The XP-endo Shaper, iRaCe, and EdgeFile systems showed a similar shaping ability. Despite the XP-endo Shaper had significantly altered the overall geometry of the root canal to a more conical shape, neither technique was capable of completely preparing the long oval-shaped canals of mandibular incisors.
本研究使用微计算机断层扫描(micro-CT)技术评估了 XP-endo Shaper(FKG Dentaire SA,La Chaux-de-Fonds,瑞士)、iRaCe(FKG Dentaire SA)和 EdgeFile(EdgeEndo,Albuquerque,NM)系统的成形能力。
使用 micro-CT 扫描(SkyScan1174v2;Bruker-microCT,Kontich,比利时)对 30 个来自下颌切牙的长椭圆形根管进行解剖匹配,并根据根管预备方案(即 XP-endo Shaper、iRaCe 和 EdgeFile 系统)将其分为 3 组(n=10)。对预备前后的配准图像进行评估,以测量体积、表面积、结构模型指数(SMI)、未触及壁、面积、周长、圆度和直径的形态计量学测量值。使用单向方差分析事后 Tukey 检验对组间数据进行比较,使用配对样本 t 检验对组内数据进行比较(α=5%)。
组内比较,预备后所有测试参数均显著增加(P<0.05)。组间未触及壁的体积(~52%)和表面积(10.8%-14.2%)或未触及壁的百分比(8.17%-9.83%)的平均增加率无统计学差异(P>0.05)。与其他两组相比,XP-endo Shaper 显著改变了根管的整体几何形状,使其更接近圆锥形(SMI=2.59)(P<0.05)。预备方案后,根尖 5mm 处根管的面积、周长、圆度以及小直径和大直径的变化在组间无差异(P>0.05)。
XP-endo Shaper、iRaCe 和 EdgeFile 系统具有相似的成形能力。尽管 XP-endo Shaper 显著改变了根管的整体几何形状,使其更接近圆锥形,但两种技术均不能完全预备下颌切牙的长椭圆形根管。