An evaluation of the Takeda UA751 automatic sphygmomanometer.

作者信息

Johnston D W, Shah D

出版信息

Behav Res Ther. 1989;27(2):203-4. doi: 10.1016/0005-7967(89)90080-6.

Abstract

The Takeda UA751, a simple automatic blood pressure measuring device, was compared with standard osciltatory measurement. There was little or no systematic difference in pressure determined by the automatic or manual method and the correlation between the two devices for both systolic and distolic pressure was satisfactory. This suggests that the Takeda UA751 is an appropriate device for patient self monitoring of blood pressure or as a measure of pressure in group outcome trials. The results essential replicate an evaluation by Steptoe and Molineux (Behav. Res. Ther. 24 223-226, 1986) of an earlier version of the Takeda (the Copal UA251) that used a different method for determining blood pressure.

摘要

将武田UA751(一种简单的自动血压测量设备)与标准振荡测量法进行了比较。自动测量法和手动测量法所测定的血压几乎没有或不存在系统性差异,并且两种设备在收缩压和舒张压方面的相关性令人满意。这表明武田UA751是一种适用于患者自我监测血压的设备,或可作为群组结果试验中的压力测量工具。这些结果实质上重复了斯特普托和莫利纽克斯(《行为研究与治疗》24卷,223 - 226页,1986年)对武田早期版本(科帕尔UA251)的评估,该早期版本使用了不同的血压测定方法。