Suppr超能文献

新型便携式电子血压计(科帕 UA251)与霍克斯利随机零点血压计的比较。

Comparison of a new portable electronic sphygmomanometer (Copal UA251) with the Hawksley random zero machine.

作者信息

Malatino L S, Brown W C

机构信息

Medical Research Council, Blood Pressure Unit, Western Infirmary, Glasgow, U.K.

出版信息

Clin Exp Hypertens A. 1988;10(4):589-96. doi: 10.3109/10641968809033911.

Abstract

The performance of the Copal Digital Sphygmomanometer UA251, a new semi-automatic blood pressure recorder, was evaluated by comparing the results obtained simultaneously and on the same arm using the standard zero-muddler sphygmomanometer. The study was performed on 67 hypertensive patients displaying a wide range of blood pressure and arm circumference. The agreement was acceptable with a mean difference, using the first set of readings of -0.45 (S.D. 2.9, range 8 to -5) and -0.95 (S.D. 2.6, range 6 to -4) mmHg for the systolic and diastolic pressures respectively. The repeatability of the Copal machine was also good; the mean difference between first and second measurements was -0.21 (S.D. 3.4, range 11 to -11) and -0.69 (S.D. 1.9, range 5 to -8) mmHg for systolic and diastolic readings respectively. Since this electronic machine has an acceptable accuracy, is readily portable, simple to use and relatively inexpensive, it seems suitable for use both in clinics and for more frequent monitoring of blood pressure throughout the day.

摘要

通过使用标准的零汞柱血压计,对同一手臂同时测量所获得的结果进行比较,对新型半自动血压记录仪科帕数字血压计UA251的性能进行了评估。该研究对67名血压范围和手臂周长差异较大的高血压患者进行。一致性是可接受的,收缩压和舒张压的第一组读数的平均差值分别为-0.45(标准差2.9,范围8至-5)和-0.95(标准差2.6,范围6至-4)mmHg。科帕血压计的重复性也很好;收缩压和舒张压读数的第一次和第二次测量之间的平均差值分别为-0.21(标准差3.4,范围11至-11)和-0.69(标准差1.9,范围5至-8)mmHg。由于这种电子血压计具有可接受的准确性、便于携带、使用简单且相对便宜,它似乎既适用于诊所,也适用于全天更频繁的血压监测。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验