Blood pressure self-measurement in upper arm and in wrist for treatment control of arterial hypertension compared to ABPM.

作者信息

Eckert S, Gleichmann S, Gleichmann U

机构信息

Kardiologische Klinik Herzzentrum, Nordrhein-Westfalen, Universitätsklinik, Ruhr-Universität Bochum, Germany.

出版信息

Z Kardiol. 1996;85 Suppl 3:109-11.

PMID:8896310
Abstract

Automatic 24 h Ambulatory Blood Pressure Monitoring (ABPM) has become an established method in the diagnosis and treatment control of arterial hypertension. Fully automatic blood pressure devices for blood pressure self-measurement (BPSM) in the upper arm and in the wrist using the oscillometric measuring technique have been available in Germany for several years. After evaluating the blood pressure devices boso Oscillomat (upper arm) and Omron HEM-601 (wrist) by means of simultaneous invasive measurements, we studied the suitability of this new method for the control of treated hypertension. Using the device Space Labs 90207, we had ABPM carried out in 100 men (aged 58 +/- 8 years) with mild to moderate hypertension. They recorded hourly blood pressure measurements in the wrist between 6 a.m. and 6 p.m., in the same arm as the blood pressure monitoring using the device Omron HEM-601, or in the other arm using the device boso Oscillomat. The values of systolic and diastolic blood pressure were measured using both methods. Since the random samples probably consisted of non-standard distribution differences, we applied the Wilcoxon matched pairs signed rank test for statistical analysis. No significant differences were recorded between the daily mean systolic (p < .07) and diastolic (p < .97) blood pressure values determined by ABPM and by frequent blood pressure self-measurement. Multiple blood pressure readings obtained using blood pressure self-measurement in the upper arm as well as in the wrist are an alternative to the established, automatic 24-h, ambulatory blood pressure measurement for the treatment control of arterial hypertension, that nightly blood pressure measurement can be omitted. It has advantage of unrestricted and desired frequent use under standardized conditions.

摘要

动态血压监测(ABPM)已成为诊断和治疗控制动脉高血压的既定方法。德国已有数年头可使用采用示波测量技术的上臂和手腕全自动血压自我测量(BPSM)设备。在通过同步侵入性测量评估了博索示波仪(上臂)和欧姆龙HEM - 601(手腕)血压设备之后,我们研究了这种新方法对已治疗高血压控制的适用性。我们使用太空实验室90207设备,对100名年龄在58±8岁之间的轻度至中度高血压男性进行了动态血压监测。他们于上午6点至下午6点期间,使用欧姆龙HEM - 601设备在与动态血压监测同一侧手臂或使用博索示波仪在另一侧手臂每小时记录一次手腕血压测量值。两种方法均测量收缩压和舒张压值。由于随机样本可能存在非标准分布差异,我们应用威尔科克森配对符号秩检验进行统计分析。动态血压监测和频繁血压自我测量所确定的每日平均收缩压(p < 0.07)和舒张压(p < 0.97)值之间未记录到显著差异。在上臂以及手腕进行血压自我测量所获得的多次血压读数,可替代既定的24小时动态血压测量用于动脉高血压的治疗控制,且可省略夜间血压测量。它具有在标准化条件下不受限制且可按需频繁使用的优点。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验