毛细管电泳系统(CAPILLARYS 2)检测血红蛋白变异体的精准度基于其迁移位置。

Precision of CAPILLARYS 2 for the Detection of Hemoglobin Variants Based on Their Migration Positions.

机构信息

Genetic Biochemistry Department, Henri Mondor Hospital, Creteil, France.

Division of Hematopathology, Mayo Clinic, Rochester, MN.

出版信息

Am J Clin Pathol. 2018 Jan 29;149(2):172-180. doi: 10.1093/ajcp/aqx148.

Abstract

OBJECTIVES

In this report, we evaluated utility of the capillary electrophoresis (CE) migration position of the CAPILLARYS 2 CE instrument.

METHODS

The precision of this x-axis number was determined on a selection of common hemoglobin (Hb) variants (Hb S, Hb C, Hb D-Punjab, Hb E, Hb Hope), and the reproducibility of this number was evaluated by comparing the results obtained by two large reference laboratories on 81 Hb variants. Additionally, the CE migration position is given for a total of 409 Hb variants.

RESULTS

The x-axis migration position showed excellent intra- and interassay precision. Comparison of Hb variants seen by both laboratories showed that 83% had a difference in migration position of 1 unit or less. Only three rare Hb variants showed a difference of more than 2 units.

CONCLUSION

In summary, the CE migration position is a reproducible value and can be used as an aid in the identification of Hb variants.

摘要

目的

本报告评估了毛细管电泳(CE)仪器毛细管电泳(CE)迁移位置的实用性。

方法

选择常见的血红蛋白(Hb)变异体(Hb S、Hb C、Hb D-Punjab、Hb E、Hb Hope)来确定此 x 轴数字的精密度,并通过比较两个大型参考实验室对 81 种 Hb 变异体的结果来评估此数字的重现性。此外,还给出了总共 409 种 Hb 变异体的 CE 迁移位置。

结果

x 轴迁移位置显示出极好的内和间实验精度。对两个实验室所见 Hb 变异体的比较表明,83%的变异体的迁移位置差异在 1 个单位或以下。只有三种罕见的 Hb 变异体的差异超过 2 个单位。

结论

总之,CE 迁移位置是一个可重复的值,可以作为鉴定 Hb 变异体的辅助手段。