非散瞳眼底摄影测量盘-黄斑中心凹夹角的可靠性和可重复性

Reliability and reproducibility of disc-foveal angle measurements by non-mydriatic fundus photography.

作者信息

Le Jeune Caroline, Chebli Fayçal, Leon Lorette, Anthoine Emmanuelle, Weber Michel, Péchereau Alain, Lebranchu Pierre

机构信息

Department of Ophthalmology, University Hospital of Nantes, Nantes, France.

Department of Ophthalmology, Docteur Nekkache Hospital, Algiers, Algeria.

出版信息

PLoS One. 2018 Jan 25;13(1):e0191007. doi: 10.1371/journal.pone.0191007. eCollection 2018.

Abstract

PURPOSE

Abnormal torsion could be associated with cyclovertical strabismus, but torsion measurements are not reliable in children. To assess an objective fundus torsion evaluation in a paediatric population, we used Non-Mydriatic Fundus photography (NMFP) in healthy and cyclovertical strabismus patients to evaluate the disc-foveal angle over time and observers.

METHODS

We used a retrospective set of NMFP including 24 A or V-pattern strabismus and 27 age-matched normal children (mean age 6.4 and 6.7 years respectively), taken during 2 distinct follow-up consultations (separated by 251 and 479 days respectively). Each disc-foveal angle measurement (from which the ocular torsion can be assessed) was performed by 5 different observers, using graphical software and based on reproducible fundus anatomical marks. Statistical analysis was performed with a multivariate ANOVA using group, time and observers as factors, in addition to intraclass coefficient correlation (ICC) to assess measurement reproducibility.

RESULTS

A significant difference of disc-foveal angle measures was observed between groups (p<0,001): 18.73° (SD = 6.42), -3,25° (SD = 5.51) and 6,89° (SD = 4,41) respectively for V-pattern, A- pattern and normal subjects. Neither observers (F = 0,2028 p = 0,9369) nor time between 1st and 2nd NMFP (F = 0,6312 p = 0,4271) seem to influence the measure of disc-foveal angle. The evaluation of disc-foveal angle was very reproducible between observers (ICC>0,97).

CONCLUSION

Abnormal amount of objective torsion could be associated with alphabet-pattern strabismus. Disc-foveal angle evaluation by NMFP in a children population appears as a non-invasive, reliable and reproducible method.

摘要

目的

异常扭转可能与旋转性斜视有关,但儿童的扭转测量并不可靠。为了评估儿科人群中客观的眼底扭转情况,我们对健康儿童和旋转性斜视患者使用非散瞳眼底摄影(NMFP)来评估不同时间及不同观察者之间的视盘-黄斑中心凹夹角。

方法

我们采用回顾性NMFP数据集,其中包括24例A或V型斜视儿童以及27例年龄匹配的正常儿童(平均年龄分别为6.4岁和6.7岁),这些数据来自两次不同的随访会诊(分别间隔251天和479天)。每次视盘-黄斑中心凹夹角测量(可据此评估眼球扭转)由5名不同的观察者进行,使用图形软件并基于可重复的眼底解剖标志。采用多因素方差分析,将组、时间和观察者作为因素进行统计分析,此外还使用组内相关系数(ICC)评估测量的可重复性。

结果

观察到组间视盘-黄斑中心凹夹角测量存在显著差异(p<0.001):V型、A 型和正常受试者的视盘-黄斑中心凹夹角分别为18.73°(标准差 = 6.42)、-3.25°(标准差 = 5.51)和6.89°(标准差 = 4.41)。观察者(F = 0.2028,p = 0.9369)以及第一次和第二次NMFP之间的时间间隔(F = 0.6312,p = 0.4271)似乎均不影响视盘-黄斑中心凹夹角的测量。观察者之间对视盘-黄斑中心凹夹角的评估具有很高的可重复性(ICC>0.97)。

结论

客观扭转量异常可能与字母型斜视有关。在儿童群体中,通过NMFP评估视盘-黄斑中心凹夹角似乎是一种非侵入性、可靠且可重复的方法。