Suppr超能文献

[连续波超声多普勒法测量血流量的实验室实验研究:II. 流量测量]

[Laboratory experimental studies on quantitative blood flow measurement with the c.w. ultrasound Doppler procedure: II. Flow measurements].

作者信息

Hildebrandt J, Kuttke P, Grossmann H

出版信息

Ultraschall Med. 1986 Apr;7(2):98-102. doi: 10.1055/s-2007-1011923.

Abstract

The reliability of Doppler blood flow measurement was examined by testing a cw Doppler flowmeter on a flow-rig. Comparing the Doppler measurements with those simultaneously obtained by means of EFM, the correlation coefficient was found to be 0.989 for continuous, and 0.985 for pulsatile streaming. With velocities up to 40 cm/sec there was a systemic overestimation of flow velocity which was attributed to the effects of the high pass filter used. Unequal distribution of intensity within the sound field is discussed. Systemic underestimation of velocities over 45 cm/sec is due to the interaction of the low pass filter. Identification of the 90 degrees reference angle using the cw Doppler technique seems to be appropriate. Within zones of turbulence caused by eccentric stenoses no reliable flow values could be determined.

摘要

通过在血流模拟装置上测试连续波多普勒流量计,对多普勒血流测量的可靠性进行了检验。将多普勒测量结果与同时通过电磁流量计获得的测量结果进行比较,发现连续血流的相关系数为0.989,脉动血流的相关系数为0.985。在流速高达40厘米/秒时,存在对流速的系统性高估,这归因于所使用的高通滤波器的影响。讨论了声场中强度分布不均的问题。流速超过45厘米/秒时的系统性低估是由于低通滤波器的相互作用。使用连续波多普勒技术识别90度参考角似乎是合适的。在由偏心狭窄引起的湍流区域内,无法确定可靠的血流值。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验