Suppr超能文献

使用质子治疗设施中的机房 CT 图像引导技术评估前列腺癌治疗的定位准确性和日剂量。

Positioning accuracy and daily dose assessment for prostate cancer treatment using in-room CT image guidance at a proton therapy facility.

机构信息

Proton Therapy Center, Fukui Prefectural Hospital, Fukui, 9108526, Japan.

Research & Development Department, The Wakasa Wan Energy Research Center, Fukui, 9140192, Japan.

出版信息

Med Phys. 2018 May;45(5):1832-1843. doi: 10.1002/mp.12858. Epub 2018 Apr 1.

Abstract

PURPOSE

To evaluate the effectiveness of CT image-guided proton radiotherapy for prostate cancer by analyzing the positioning uncertainty and assessing daily dose change due to anatomical variations.

MATERIALS AND METHODS

Patients with prostate cancer were treated by opposed lateral proton beams based on a passive scattering method using an in-room CT image-guided system. The system employs a single couch for both CT scanning and beam delivery. The patient was positioned by matching the boundary between the prostate and the rectum's anterior region identified in the CT images to the corresponding boundary in the simulator images after bone matching. We acquired orthogonal kV x-ray images after couch movement and confirmed the body position by referring to the bony structure prior to treatment. In offline analyses, we contoured the targeted anatomical structures on 375 sets of daily in-room CT images for 10 patients. The uncertainty of the image-matching procedure was evaluated using the prostate contours and actual couch corrections. We also performed dose calculations using the same set of CT images, and evaluated daily change of dose-volume histograms (DVHs) to compare the effectiveness of the treatment using prostate matching to the bone-matching procedure.

RESULTS

The isocenter shifts by prostate matching after bone matching were 0.5 ± 1.8 and -0.8 ± 2.6 mm along the superior-inferior (SI) and anterior-posterior (AP) directions, respectively. The body movement errors (σ) after couch movement were 0.7, 0.5, and 0.3 mm along the lateral, SI and AP direction, respectively, for 30 patients. The estimated errors (σ) in the prostate matching were 1.0 and 1.3 mm, and, in conjunction with the movement errors, the total positioning uncertainty was estimated to be 1.0 and 1.4 mm along the SI and AP directions, respectively. Daily DVH analyses showed that in the prostate matching, 98.7% and 86.1% of the total 375 irradiations maintained a dose condition of V  > 95% for the prostate and a dose constraint of V  < 18% for the rectum, whereas 90.4% and 66.1% of the total irradiations did so when bone matching was used. The dose constraint of the rectum and dose coverage of the prostate were better maintained by prostate matching than bone matching (P < 0.001). The daily variation in the dose to the seminal vesicles (SVs) was large, and only 40% of the total irradiations maintained the initial planned values of V for high-risk treatment. Nevertheless, the deviations from the original value were -4 ± 7% and -5 ± 11% in the prostate and bone matching, respectively, and a better dose coverage of the SV was achieved by the prostate matching.

CONCLUSION

The correction of repositioning along the AP and SI direction from conventional bone matching in CT image-guided proton therapy was found to be effective to maintain the dose constraint of the rectum and the dose coverage of the prostate. This work indicated that prostate cancer treatment by prostate matching using CT image guidance may be effective to reduce the rectal complications and achieve better tumor control of the prostate. However, an adaptive approach is desirable to maintain better dose coverage of the SVs.

摘要

目的

通过分析定位不确定性并评估因解剖结构变化导致的每日剂量变化,评估 CT 图像引导质子放射治疗前列腺癌的效果。

材料和方法

10 名前列腺癌患者采用被动散射方法基于对侧横向质子束进行治疗,使用室内 CT 图像引导系统。该系统使用单个治疗床同时进行 CT 扫描和束流传输。患者通过将 CT 图像中识别的前列腺和直肠前区域的边界与模拟器图像中相应的边界匹配来定位。在治疗前,我们在治疗床移动后获取正交千伏 X 射线图像,并参考骨性结构确认患者体位。在离线分析中,我们对 10 名患者的 375 组每日室内 CT 图像勾画了目标解剖结构。使用前列腺轮廓和实际治疗床修正评估图像匹配过程的不确定性。我们还使用相同的 CT 图像进行剂量计算,并评估剂量体积直方图 (DVH) 的每日变化,以比较使用前列腺匹配和骨匹配程序治疗的效果。

结果

在骨匹配后进行前列腺匹配时,等中心沿上下(SI)和前后(AP)方向的偏移分别为 0.5 ± 1.8 和-0.8 ± 2.6 毫米。30 名患者治疗床移动后的体动误差(σ)分别为 0.7、0.5 和 0.3 毫米,沿侧向、SI 和 AP 方向。前列腺匹配的估计误差(σ)为 1.0 和 1.3 毫米,加上运动误差,SI 和 AP 方向的总定位不确定性分别估计为 1.0 和 1.4 毫米。每日 DVH 分析显示,在前列腺匹配中,98.7%和 86.1%的总 375 次照射保持了前列腺 V  > 95%的剂量条件和直肠 V  < 18%的剂量限制,而当使用骨匹配时,90.4%和 66.1%的总照射保持了这一条件。与骨匹配相比,前列腺匹配更好地保持了直肠的剂量限制和前列腺的剂量覆盖(P < 0.001)。精囊(SVs)的剂量变化较大,只有 40%的总照射保持了高危治疗的初始计划 V 值。然而,在前列腺和骨匹配中,偏离原始值分别为-4 ± 7%和-5 ± 11%,并且前列腺匹配实现了更好的 SV 剂量覆盖。

结论

在 CT 图像引导质子治疗中,从常规的骨匹配进行沿 AP 和 SI 方向的重新定位校正被发现可以有效地维持直肠的剂量限制和前列腺的剂量覆盖。这项工作表明,使用 CT 图像引导的前列腺匹配治疗前列腺癌可能有助于降低直肠并发症的风险,并实现更好的前列腺肿瘤控制。然而,为了更好地保持 SVs 的剂量覆盖,需要采用自适应方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验