Despite the development of various methods of intraocular pressure (IOP) measurement, Goldmann applanation tonometry (GAT) is still the most popular. The measurement using GAT depends on the biomechanical properties of the cornea, such as the thickness, the radius of curvature, as well as the amount of the fluorescein used.
The aim of the study was to compare IOP values measured by GAT with those measured by applanation resonance tonometry (ART).
A total of 47 patients (94 eyes), including 28 patients with primary open-angle glaucoma (POAG) and 19 subjects from the control group, were examined at the Glaucoma Outpatient Clinic of the Department and Clinic of Ophthalmology at Wroclaw Medical University (Poland). The measurements of IOP were performed using GAT and a handheld version of ART. Also, the central corneal thickness (CCT) of all patients was measured.
The study showed that the IOP values measured by both tonometers were comparable, but ARTacquired values were higher than GAT-obtained values both in the glaucomatous group and in the control group. CCT had little impact on mean IOP difference between GATand ART-obtained values.
Applanation resonance tonometry is a precise method of IOP measurement and is less affected by biomechanical properties of the cornea than GAT. Our results show that ART is a new, promising, comfortable for both patients and doctors method of IOP measurement, which, in the future, can replace GAT.
尽管眼压(IOP)测量方法多种多样,但Goldmann压平眼压计(GAT)仍是最常用的。使用GAT进行测量取决于角膜的生物力学特性,如厚度、曲率半径以及所用荧光素的量。
本研究旨在比较GAT测量的眼压值与压平式共振眼压计(ART)测量的眼压值。
在波兰弗罗茨瓦夫医科大学眼科系和诊所的青光眼门诊,对47例患者(94只眼)进行了检查,其中包括28例原发性开角型青光眼(POAG)患者和19例对照组受试者。使用GAT和手持式ART进行眼压测量。此外,还测量了所有患者的中央角膜厚度(CCT)。
研究表明,两种眼压计测量的眼压值具有可比性,但在青光眼组和对照组中,ART测得的值均高于GAT测得的值。CCT对GAT和ART测得值之间的平均眼压差异影响不大。
压平式共振眼压计是一种精确的眼压测量方法,与GAT相比,受角膜生物力学特性的影响较小。我们的结果表明,ART是一种新型、有前景且患者和医生都适用的眼压测量方法,未来有望取代GAT。