Omer Khaled, Fox Kathryn, Palermo Deniel, Boyle Liam, Youngson Callum
Department of Restorative Dentistry, Liverpool Dental School, Liverpool, UK.
BDJ Open. 2015 Oct 23;1:15001. doi: 10.1038/bdjopen.2015.1. eCollection 2015.
OBJECTIVES/AIMS: To determine whether a clinically available resonance frequency analysis (RFA) device (Osstell Mentor), designed to assess implant integration, could identify a single uncemented retainer on fixed-fixed bridges, .
All-metal fixed-fixed bridges were cemented to acrylic tooth analogue abutments with simulated periodontal ligaments. Dental stone bases provided 100 or 50% 'bone support' groups ( = 50 in each). Control groups had both retainers cemented, whereas the test groups had the 'premolar' uncemented, mimicking clinical failure. A RFA device was used to measure bridge stability, expressed as a modified Implant Stability Quotient (Bridge Stability Quotient-BSQ) from a Smartpeg temporarily affixed to the bridge via composite.
The BSQ recorded at the premolar site in both 100 and 50% support models demonstrated highly statistically significant differences ( <0.003) between the control and test groups. Sensitivity and specificity, area under the curve (receiver operating characteristic), analyses showed moderate test accuracy (0.735) for the 100% support group and good accuracy (0.96) for the 50% support group.
The investigation suggests that RFA measurements were able to identify, reliably and non-destructively, , fixed-fixed bridges where the anterior retainer was uncemented. Further clinical research is required to determine whether this technique may allow early diagnosis of failing bridgework.
确定一种旨在评估种植体整合情况的临床可用共振频率分析(RFA)设备(Osstell Mentor)能否识别固定-固定桥中单个未粘结的固位体。
全金属固定-固定桥粘结到带有模拟牙周韧带的丙烯酸牙齿模型基台上。石膏底座分为提供100%或50%“骨支持”的组(每组n = 50)。对照组的两个固位体均粘结,而试验组的“前磨牙”固位体未粘结,模拟临床失败情况。使用RFA设备测量桥体稳定性,通过临时用复合材料固定在桥体上的Smartpeg以改良种植体稳定性商数(桥体稳定性商数-BSQ)表示。
在100%和50%支持模型的前磨牙部位记录的BSQ显示,对照组和试验组之间存在高度统计学显著差异(P <0.003)。敏感性和特异性、曲线下面积(受试者工作特征)分析表明,100%支持组的测试准确性中等(0.735),50%支持组的准确性良好(0.96)。
该研究表明,RFA测量能够可靠且无损地识别前固位体未粘结的固定-固定桥。需要进一步的临床研究来确定该技术是否可实现对失败桥体修复的早期诊断。