Fernández Joaquín, Rodríguez-Vallejo Manuel, Martínez Javier, Tauste Ana, García-Montesinos Javier, Piñero David P
Department of Ophthalmology (Qvision), Vithas Virgen del Mar Hospital, 04120, Almería, Spain.
Department of Ophthalmology, Torrecárdenas Hospital Complex, 04009, Almería, Spain.
Graefes Arch Clin Exp Ophthalmol. 2018 Aug;256(8):1535-1541. doi: 10.1007/s00417-018-3986-9. Epub 2018 Apr 18.
To assess the agreement and repeatability of two objective systems for measuring the tear film stability.
Retrospective analysis of the tear film stability of 99 healthy right eyes measured with a videokeratoscope (VK) and the Optical Quality Analysis System (OQAS, Visiometrics). Two consecutive measures were taken with both systems, with an interval of 10 min between them. Variables included in the study were first and mean non-invasive break-up times (NIBUT and MNIBUT) measured with VK, and mean and standard deviation of the optical scattering index (OSIm and OSIsd) measured with OQAS. The agreement and repeatability of grading scales provided by both devices were also evaluated using the Cohen's k with quadratic weights. The Ocular Surface Disease index (OSDI) questionnaire was also passed out to all subjects. Correlations and associations between subjective and objective metrics were analyzed.
Significant differences were found between consecutive measurements of NIBUT (p = 0.04) and MNIBUT (p = 0.01), but not for OSIm (p = 0.11) and OSIsd (p = 0.50). Grading scales resulted in fair (k = 0.20) or poor agreement (k = 0.04) between systems depending if the first or second trial was considered. The repeatability of the grading scale was good for OQAS (k = 0.59) and fair for VK (k = 0.37). No significant correlations or associations were found between OSDI and any of the metrics obtained with both devices (p ≥ 0.36).
The two devices evaluated cannot be used interchangeably for the assessment of tear film stability. Good intrasession repeatability was obtained for tear film grading of the OQAS whereas it was fair for VK.
评估两种测量泪膜稳定性的客观系统的一致性和可重复性。
对99只健康右眼的泪膜稳定性进行回顾性分析,使用角膜地形图仪(VK)和光学质量分析系统(OQAS,Visiometrics公司)进行测量。两种系统均进行连续两次测量,两次测量间隔10分钟。研究中纳入的变量包括用VK测量的首次和平均非侵入性泪膜破裂时间(NIBUT和MNIBUT),以及用OQAS测量的光学散射指数的平均值和标准差(OSIm和OSIsd)。还使用二次加权的科恩kappa系数评估了两种设备提供的分级量表的一致性和可重复性。还向所有受试者发放了眼表疾病指数(OSDI)问卷。分析了主观和客观指标之间的相关性和关联性。
连续测量NIBUT(p = 0.04)和MNIBUT(p = 0.01)之间存在显著差异,但OSIm(p = 0.11)和OSIsd(p = 0.50)之间无显著差异。根据考虑的是第一次还是第二次试验,分级量表在系统之间的一致性为中等(k = 0.20)或较差(k = 0.04)。分级量表的可重复性对于OQAS良好(k = 0.59),对于VK中等(k = 0.37)。在OSDI与两种设备获得的任何指标之间均未发现显著的相关性或关联性(p≥0.36)。
所评估的两种设备不能互换用于评估泪膜稳定性。OQAS的泪膜分级具有良好的测量期间可重复性,而VK的可重复性中等。