Tian Lei, Qu Jing-Hao, Zhang Xiao-Yu, Sun Xu-Guang
Beijing Institute of Ophthalmology, Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University and Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing 100730, China.
J Ophthalmol. 2016;2016:8013621. doi: 10.1155/2016/8013621. Epub 2016 Apr 12.
Purpose. To determine the intraexaminer repeatability and interexaminer reproducibility of tear meniscus height (TMH) and noninvasive Keratograph tear breakup time (NIKBUT) measurements obtained with the Keratograph 5M (K5M) in a sample of healthy and dry eye populations. Methods. Forty-two patients with dry eye disease (DED group) and 42 healthy subjects (healthy group) were recruited in this prospective study. In all subjects, each eye received 3 consecutive measurements using the K5M for the TMH and NIKBUTs (NIKBUT-first and NIKBUT-average). And then a different examiner repeated the measurements. The repeatability and reproducibility of measurements were assessed by the coefficient of variation (CV) and intraclass correlation coefficient (ICC). Results. The repeatability and reproducibility of TMH and NIKBUTs were good in both DED and healthy groups (CV% ≤ 26.1% and ICC ≥ 0.75 for all measurements). Patients with DED showed better intraexaminer repeatability for NIKBUTs, but worse for TMH than healthy subjects. Average TMH, NIKBUT-first, and NIKBUT-average were significantly lower in DED group than in healthy group (all P values < 0.05). Conclusions. Measurements of TMH and NIKBUTs obtained with the K5M may provide a simple, noninvasive screening test for dry eye with acceptable repeatability and reproducibility. The NIKBUTs were more reliable, but TMH was less reliable in patients with DED.
目的。确定在健康人群和干眼人群样本中,使用角膜地形图仪5M(K5M)获得的泪膜半月板高度(TMH)和无创角膜地形图仪泪膜破裂时间(NIKBUT)测量值的检查者内重复性和检查者间再现性。方法。在这项前瞻性研究中招募了42例干眼疾病患者(干眼组)和42名健康受试者(健康组)。在所有受试者中,每只眼睛使用K5M对TMH和NIKBUT(首次NIKBUT和平均NIKBUT)进行连续3次测量。然后由另一名检查者重复测量。通过变异系数(CV)和组内相关系数(ICC)评估测量的重复性和再现性。结果。在干眼组和健康组中,TMH和NIKBUT的重复性和再现性均良好(所有测量的CV%≤26.1%且ICC≥0.75)。干眼患者的NIKBUT检查者内重复性较好,但TMH的检查者内重复性比健康受试者差。干眼组的平均TMH、首次NIKBUT和平均NIKBUT均显著低于健康组(所有P值<0.05)。结论。使用K5M获得的TMH和NIKBUT测量值可为干眼提供一种具有可接受的重复性和再现性的简单、无创筛查测试。在干眼患者中,NIKBUT更可靠,但TMH不太可靠。