Quantitative diffusion MRI using reduced field-of-view and multi-shot acquisition techniques: Validation in phantoms and prostate imaging.

作者信息

Zhang Yuxin, Holmes James, Rabanillo Iñaki, Guidon Arnaud, Wells Shane, Hernando Diego

机构信息

Department of Medical Physics, University of Wisconsin Madison, Madison, WI, United States; Department of Radiology, School of Medicine and Public Health, University of Wisconsin Madison, Madison, WI, United States.

Department of Radiology, School of Medicine and Public Health, University of Wisconsin Madison, Madison, WI, United States.

出版信息

Magn Reson Imaging. 2018 Sep;51:173-181. doi: 10.1016/j.mri.2018.04.006. Epub 2018 Apr 17.

Abstract

PURPOSE

To evaluate the reproducibility of quantitative diffusion measurements obtained with reduced Field of View (rFOV) and Multi-shot EPI (msEPI) acquisitions, using single-shot EPI (ssEPI) as a reference.

METHODS

Diffusion phantom experiments, and prostate diffusion-weighted imaging in healthy volunteers and patients with known or suspected prostate cancer were performed across the three different sequences. Quantitative diffusion measurements of apparent diffusion coefficient, and diffusion kurtosis parameters (healthy volunteers), were obtained and compared across diffusion sequences (rFOV, msEPI, and ssEPI). Other possible confounding factors like b-value combinations and acquisition parameters were also investigated.

RESULTS

Both msEPI and rFOV have shown reproducible quantitative diffusion measurements relative to ssEPI; no significant difference in ADC was observed across pulse sequences in the standard diffusion phantom (p = 0.156), healthy volunteers (p ≥ 0.12) or patients (p ≥ 0.26). The ADC values within the non-cancerous central gland and peripheral zone of patients were 1.29 ± 0.17 × 10 mm/s and 1.74 ± 0.23 × 10 mm/s respectively. However, differences in quantitative diffusion parameters were observed across different number of averages for rFOV, and across b-value groups and diffusion models for all the three sequences.

CONCLUSION

Both rFOV and msEPI have the potential to provide high image quality with reproducible quantitative diffusion measurements in prostate diffusion MRI.

摘要

目的

以单次激发回波平面成像(ssEPI)作为参考,评估采用缩小视野(rFOV)和多次激发回波平面成像(msEPI)采集获得的定量扩散测量值的可重复性。

方法

对健康志愿者以及已知或疑似前列腺癌患者进行了扩散体模实验和前列腺扩散加权成像,采用三种不同序列。获取并比较了表观扩散系数的定量扩散测量值以及扩散峰度参数(健康志愿者)在不同扩散序列(rFOV、msEPI和ssEPI)之间的差异。还研究了其他可能的混杂因素,如b值组合和采集参数。

结果

相对于ssEPI,msEPI和rFOV均显示出可重复的定量扩散测量值;在标准扩散体模(p = 0.156)、健康志愿者(p≥0.12)或患者(p≥0.26)中,各脉冲序列间的表观扩散系数未观察到显著差异。患者非癌性中央腺体和外周带的表观扩散系数值分别为1.29±0.17×10⁻³mm²/s和1.74±0.23×10⁻³mm²/s。然而,在rFOV不同平均次数之间,以及在所有三个序列的不同b值组和扩散模型之间,观察到了定量扩散参数的差异。

结论

rFOV和msEPI都有潜力在前列腺扩散磁共振成像中提供高质量图像以及可重复的定量扩散测量值。

相似文献

7
Study of the reduced field-of-view diffusion-weighted imaging of the breast.乳腺缩小视野扩散加权成像的研究
Clin Breast Cancer. 2014 Aug;14(4):265-71. doi: 10.1016/j.clbc.2013.12.001. Epub 2013 Dec 27.

引用本文的文献

5
Quantitative diffusion MRI of the abdomen and pelvis.腹部和盆腔的定量扩散 MRI。
Med Phys. 2022 Apr;49(4):2774-2793. doi: 10.1002/mp.15246. Epub 2021 Oct 8.

本文引用的文献

4
Cancer statistics, 2016.癌症统计数据,2016 年。
CA Cancer J Clin. 2016 Jan-Feb;66(1):7-30. doi: 10.3322/caac.21332. Epub 2016 Jan 7.
8
Understanding Bland Altman analysis.理解布兰德-奥特曼分析。
Biochem Med (Zagreb). 2015 Jun 5;25(2):141-51. doi: 10.11613/BM.2015.015. eCollection 2015.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验