Zhang Yuxin, Holmes James, Rabanillo Iñaki, Guidon Arnaud, Wells Shane, Hernando Diego
Department of Medical Physics, University of Wisconsin Madison, Madison, WI, United States; Department of Radiology, School of Medicine and Public Health, University of Wisconsin Madison, Madison, WI, United States.
Department of Radiology, School of Medicine and Public Health, University of Wisconsin Madison, Madison, WI, United States.
Magn Reson Imaging. 2018 Sep;51:173-181. doi: 10.1016/j.mri.2018.04.006. Epub 2018 Apr 17.
To evaluate the reproducibility of quantitative diffusion measurements obtained with reduced Field of View (rFOV) and Multi-shot EPI (msEPI) acquisitions, using single-shot EPI (ssEPI) as a reference.
Diffusion phantom experiments, and prostate diffusion-weighted imaging in healthy volunteers and patients with known or suspected prostate cancer were performed across the three different sequences. Quantitative diffusion measurements of apparent diffusion coefficient, and diffusion kurtosis parameters (healthy volunteers), were obtained and compared across diffusion sequences (rFOV, msEPI, and ssEPI). Other possible confounding factors like b-value combinations and acquisition parameters were also investigated.
Both msEPI and rFOV have shown reproducible quantitative diffusion measurements relative to ssEPI; no significant difference in ADC was observed across pulse sequences in the standard diffusion phantom (p = 0.156), healthy volunteers (p ≥ 0.12) or patients (p ≥ 0.26). The ADC values within the non-cancerous central gland and peripheral zone of patients were 1.29 ± 0.17 × 10 mm/s and 1.74 ± 0.23 × 10 mm/s respectively. However, differences in quantitative diffusion parameters were observed across different number of averages for rFOV, and across b-value groups and diffusion models for all the three sequences.
Both rFOV and msEPI have the potential to provide high image quality with reproducible quantitative diffusion measurements in prostate diffusion MRI.
以单次激发回波平面成像(ssEPI)作为参考,评估采用缩小视野(rFOV)和多次激发回波平面成像(msEPI)采集获得的定量扩散测量值的可重复性。
对健康志愿者以及已知或疑似前列腺癌患者进行了扩散体模实验和前列腺扩散加权成像,采用三种不同序列。获取并比较了表观扩散系数的定量扩散测量值以及扩散峰度参数(健康志愿者)在不同扩散序列(rFOV、msEPI和ssEPI)之间的差异。还研究了其他可能的混杂因素,如b值组合和采集参数。
相对于ssEPI,msEPI和rFOV均显示出可重复的定量扩散测量值;在标准扩散体模(p = 0.156)、健康志愿者(p≥0.12)或患者(p≥0.26)中,各脉冲序列间的表观扩散系数未观察到显著差异。患者非癌性中央腺体和外周带的表观扩散系数值分别为1.29±0.17×10⁻³mm²/s和1.74±0.23×10⁻³mm²/s。然而,在rFOV不同平均次数之间,以及在所有三个序列的不同b值组和扩散模型之间,观察到了定量扩散参数的差异。
rFOV和msEPI都有潜力在前列腺扩散磁共振成像中提供高质量图像以及可重复的定量扩散测量值。