Department of Anthropology, University of Pittsburgh, Pittsburgh, Pennsylvania 15260.
Center for Craniofacial and Dental Genetics, Department of Oral Biology, University of Pittsburgh, Pittsburgh, Pennsylvania 15219.
Am J Hum Biol. 2018 Jul;30(4):e23133. doi: 10.1002/ajhb.23133. Epub 2018 May 9.
We compared digital 3D stereophotogrammetry to more traditional measurement methods (direct anthropometry and 2D scanning) to capture digit lengths and ratios.
The length of the second and fourth digits was measured by each method and the second-to-fourth ratio was calculated. For each digit measurement, intraobserver agreement was calculated for each of the three collection methods. Further, measurements from the three methods were compared directly to one another. Agreement statistics included the intraclass correlation coefficient (ICC) and technical error of measurement (TEM).
Intraobserver agreement statistics for the digit length measurements were high for all three methods; ICC values exceeded 0.97 and TEM values were below 1 mm. For digit ratio, intraobserver agreement was also acceptable for all methods, with direct anthropometry exhibiting lower agreement (ICC = 0.87) compared to indirect methods. For the comparison across methods, the overall agreement was high for digit length measurements (ICC values ranging from 0.93 to 0.98; TEM values below 2 mm). For digit ratios, high agreement was observed between the two indirect methods (ICC = 0.93), whereas indirect methods showed lower agreement when compared to direct anthropometry (ICC < 0.75).
Digit measurements and derived ratios from 3D stereophotogrammetry showed high intraobserver agreement (similar to more traditional methods) suggesting that landmarks could be placed reliably on 3D hand surface images. While digit length measurements were found to be comparable across all three methods, ratios derived from direct anthropometry tended to be higher than those calculated indirectly from 2D or 3D images.
我们比较了数字 3D 体视摄影与更传统的测量方法(直接人体测量和 2D 扫描),以获取手指长度和比例。
使用每种方法测量第二和第四指的长度,并计算第二至第四指的比例。对于每个手指测量,计算了三种采集方法的每个方法的观察者内一致性。此外,还直接比较了三种方法的测量值。一致性统计数据包括组内相关系数(ICC)和测量技术误差(TEM)。
三种方法的手指长度测量的观察者内一致性统计数据均较高;ICC 值均超过 0.97,TEM 值均低于 1mm。对于手指比例,所有方法的观察者内一致性也可以接受,与间接方法相比,直接人体测量的一致性较低(ICC=0.87)。对于方法之间的比较,手指长度测量的总体一致性较高(ICC 值范围为 0.93 至 0.98;TEM 值低于 2mm)。对于手指比例,两种间接方法之间观察到高度一致(ICC=0.93),而与直接人体测量相比,间接方法的一致性较低(ICC<0.75)。
3D 体视摄影的手指测量值和衍生比例具有较高的观察者内一致性(与更传统的方法相似),这表明可以在 3D 手部表面图像上可靠地标定地标。虽然发现所有三种方法的手指长度测量值都可以比较,但直接人体测量得出的比例往往高于从 2D 或 3D 图像间接计算得出的比例。