He Yong, Shiu Yan-Ting, Pike Daniel B, Roy-Chaudhury Prabir, Cheung Alfred K, Berceli Scott A
Department of Surgery, University of Florida, Gainesville, Fla; Malcom Randall VA Medical Center, Gainesville, Fla.
Division of Nephrology & Hypertension, University of Utah, Salt Lake City, Utah.
J Vasc Surg. 2018 Dec;68(6):1848-1857.e2. doi: 10.1016/j.jvs.2018.02.043. Epub 2018 May 18.
The objective of this study was to compare blood flow rates measured by Doppler ultrasound (DUS) and phase-contrast magnetic resonance imaging (MRI) in patients having a hemodialysis arteriovenous fistula (AVF) and to identify scenarios in which there was significant discordance between these two approaches.
Blood flow rates in the proximal artery (PA) and draining vein (DV) of newly created upper extremity AVFs were measured and compared using DUS and phase-contrast MRI at 1 day, 6 weeks, and 6 months postoperatively.
Blood flow rates in the PA measured by DUS (1155 ± 907 mL/min, mean ± standard deviation) and by MRI (1170 ± 657 mL/min) were not statistically different (P = .812) based on 78 data pairs from 49 patients. DV DUS flow (1277 ± 995 mL/min) and MRI flow (1130 ± 655 mL/min) were also not statistically different (P = .071) based on 64 data pairs. In both PA and DV, the two methods substantially agreed with each other (Cohen κ: PA, 0.66; DV, 0.67) when flow rates were put into four clinically relevant categories (<300, 300-599, 600-1499, and ≥1500 mL/min). The Bland-Altman analyses of DUS and MRI flow identified six and four outliers for PA and DV, respectively. Seven outliers had higher DUS than MRI flow, with all DUS scan sites having a large lumen or significant local curvature; the other three had lower DUS flow, partly due to an underestimation of lumen diameter by DUS.
DUS and MRI flow rates are generally comparable in both PA and DV. When DUS is used for flow measurements, careful attention to accurate lumen diameter measurements is needed and scan sites with marked curvature should be avoided. Our result may improve the accuracy of DUS-measured AVF blood flow rate.
本研究的目的是比较采用多普勒超声(DUS)和相位对比磁共振成像(MRI)测量血液透析动静脉内瘘(AVF)患者的血流速率,并确定这两种方法存在显著差异的情况。
在术后1天、6周和6个月时,使用DUS和相位对比MRI测量并比较新建立的上肢AVF的近端动脉(PA)和引流静脉(DV)的血流速率。
基于49例患者的78对数据,DUS测量的PA血流速率(1155±907 mL/min,平均值±标准差)与MRI测量的血流速率(1170±657 mL/min)无统计学差异(P = 0.812)。基于64对数据,DV的DUS血流(1277±995 mL/min)和MRI血流(1130±655 mL/min)也无统计学差异(P = 0.071)。当将PA和DV的血流速率分为四个临床相关类别(<300、300 - 599、600 - 1499和≥1500 mL/min)时,两种方法在PA和DV中彼此基本一致(Cohen κ:PA为0.66;DV为0.67)。DUS和MRI血流的Bland - Altman分析分别确定了PA和DV的6个和4个异常值。7个异常值的DUS血流高于MRI血流,所有DUS扫描部位的管腔较大或存在明显的局部弯曲;另外3个异常值的DUS血流较低,部分原因是DUS对管腔直径的估计偏低。
DUS和MRI测量的PA和DV血流速率总体上具有可比性。当使用DUS进行血流测量时,需要注意准确测量管腔直径,并应避免扫描具有明显弯曲的部位。我们的结果可能会提高DUS测量AVF血流速率的准确性。