骨科植入物的金属伪影减少:颅内金属植入物患者的脑 CT 血管造影。

Metal Artifact Reduction for Orthopedic Implants: Brain CT Angiography in Patients with Intracranial Metallic Implants.

机构信息

Department of Radiology, Seoul National University Bundang Hospital, Seongnam, Korea.

Department of Radiology, Seoul National University College of Medicine, Seoul, Korea.

出版信息

J Korean Med Sci. 2018 May 2;33(21):e158. doi: 10.3346/jkms.2018.33.e158. eCollection 2018 May 21.

Abstract

BACKGROUND

The purpose of this study was to qualitatively and quantitatively evaluate the effects of a metal artifact reduction for orthopedic implants (O-MAR) for brain computed tomographic angiography (CTA) in patients with aneurysm clips and coils.

METHODS

The study included 36 consecutive patients with 47 intracranial metallic implants (42 aneurysm clips, 5 coils) who underwent brain CTA. The computed tomographic images with and without the O-MAR were independently reviewed both quantitatively and qualitatively by two reviewers. For quantitative analysis, image noises near the metallic implants of non-O-MAR and O-MAR images were compared. For qualitative analysis, image quality improvement and the presence of new streak artifacts were assessed.

RESULTS

Image noise was significantly reduced near metallic implants ( < 0.01). Improvement of implant-induced streak artifacts was observed in eight objects (17.0%). However, streak artifacts were aggravated in 11 objects (23.4%), and adjacent vessel depiction was worsened in eight objects (17.0%). In addition, new O-MAR-related streak artifacts were observed in 32 objects (68.1%). New streak artifacts were more prevalent in cases with overlapping metallic implants on the same axial plane than in those without ( = 0.018). Qualitative assessment revealed that the overall image quality was not significantly improved in O-MAR images.

CONCLUSION

In conclusion, the use of the O-MAR in patients with metallic implants significantly reduces image noise. However, the degree of the streak artifacts and surrounding vessel depiction were not significantly improved on O-MAR images.

摘要

背景

本研究旨在定性和定量评估金属伪影减少技术(O-MAR)对颅内带夹动脉瘤和植入线圈的患者脑 CT 血管造影(CTA)的影响。

方法

本研究纳入了 36 例 47 个颅内金属植入物(42 个动脉瘤夹,5 个线圈)的连续患者,所有患者均接受了脑 CTA 检查。由两位观察者独立对未应用 O-MAR 技术和应用 O-MAR 技术的图像进行定性和定量评估。在定量分析中,比较了未应用 O-MAR 技术和应用 O-MAR 技术的图像中金属植入物附近的图像噪声。在定性分析中,评估了图像质量的改善和新出现的条纹伪影的情况。

结果

金属植入物附近的图像噪声显著降低(<0.01)。8 个物体(17.0%)的植入物引起的条纹伪影得到改善。然而,11 个物体(23.4%)的条纹伪影加重,8 个物体(17.0%)的相邻血管显示变差。此外,在 32 个物体(68.1%)中观察到新的 O-MAR 相关条纹伪影。在同一轴向平面上有重叠金属植入物的病例中,新的条纹伪影比没有重叠金属植入物的病例更常见(=0.018)。定性评估显示,O-MAR 图像的整体图像质量没有显著改善。

结论

总之,在金属植入物患者中使用 O-MAR 技术可显著降低图像噪声,但条纹伪影和周围血管显示的程度在 O-MAR 图像上并未显著改善。