Suppr超能文献

经睫状体平坦部玻璃体切割术后混浊亲水丙烯酸人工晶状体的材料分析与光学质量评估。

Material Analysis and Optical Quality Assessment of Opacified Hydrophilic Acrylic Intraocular Lenses After Pars Plana Vitrectomy.

机构信息

David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany.

David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany.

出版信息

Am J Ophthalmol. 2018 Sep;193:10-19. doi: 10.1016/j.ajo.2018.06.002. Epub 2018 Jun 8.

Abstract

PURPOSE

Explanted hydrophilic intraocular lenses (IOLs) with clinically significant opacification after pars plana vitrectomy (PPV) were assessed for material change and optical quality, in an in vitro laboratory study.

DESIGN

Retrospective observational case series.

METHODS

Ten opacified IOLs after PPV with intraocular gas injection were analyzed in a laboratory setting. Analyses included evaluation of patients' medical history, optical quality assessment, light microscopy, histologic staining, scanning electron microscopy, and energy dispersive x-ray spectroscopy.

RESULTS

In all 10 IOLs a thin layer of calcium phosphate that had accumulated underneath either the anterior or posterior optical surface in a central circular area of the IOL optic caused the opacification. The calcifications lead to deterioration of the modulation transfer function (MTF) across all spatial frequencies.

CONCLUSION

PPV with instillation of gas into a pseudophakic eye with an acrylic hydrophilic lens seems to increase the risk for secondary calcification irrespective of the manufacturer. In these cases, IOL exchange is the only treatment option available. Since IOL exchange is associated with a high intraoperative complication rate, our results suggest to consider the risk of IOL calcification when implanting hydrophilic acrylic IOLs.

摘要

目的

对行巩膜平坦部玻璃体切除术(PPV)后出现临床显著混浊的亲水性眼内晶状体(IOL)进行材料变化和光学质量评估,这是一项体外实验室研究。

设计

回顾性观察性病例系列。

方法

在实验室环境中分析了 10 例经 PPV 联合眼内气体注射后混浊的 IOL。分析包括评估患者的病史、光学质量评估、光学显微镜检查、组织学染色、扫描电子显微镜检查和能量色散 X 射线光谱分析。

结果

在所有 10 个 IOL 中,在 IOL 光学部的前表面或后表面下的中央圆形区域积聚的薄的磷酸钙层导致了混浊。这些钙化导致调制传递函数(MTF)在所有空间频率下都恶化。

结论

在眼后房型丙烯酸亲水晶状体的眼内注入气体的 PPV 似乎会增加继发性钙化的风险,而与制造商无关。在这些情况下,IOL 置换是唯一可用的治疗选择。由于 IOL 置换与术中并发症发生率高相关,因此我们的结果表明,在植入亲水丙烯酸 IOL 时应考虑 IOL 钙化的风险。