To investigate the efficiency of spectral computed tomography (CT) optimal monochromatic images in improving imaging quality of liver vessels.
The imaging data of 35 patients with abdominal CT angiography were retrospectively analyzed. Hepatic arteries, portal veins, and hepatic veins were reconstructed with mixed energy (quality check, QC), 70 keV and optimal monochromatic mode. Comparative parameters were analyzed including CT value, image noise (IN), contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and subjective qualitative analysis.
The optimal monochromatic value for assessment of the common hepatic artery, portal vein, and hepatic vein ranged between 49 keV and 53 keV, with a mean of 51 keV. There were statistically significant differences ( < 0.001) among the optimal monochromatic, 70 keV and QC images with regards to the hepatic vascular CT value, IN, CNR, SNR, and subjective qualitative score. CNR of the common hepatic artery in the optimal monochromatic, 70 keV and QC groups was 24.6 ± 10.9, 18.1 ± 8.3, and 11.6 ± 4.6, respectively ( < 0.001) with subjective scores of 4.7 ± 0.2, 4.0 ± 0.3, and 3.6 ± 0.4, respectively ( < 0.001). CNR of the hepatic portal vein was 6.9 ± 2.7, 4.3 ± 1.9, and 3.0 ± 2.1, respectively ( < 0.001) with subjective scores of 4.5 ± 0.3, 3.9 ± 0.4, and 3.3 ± 0.3, respectively ( < 0.001). CNR of the hepatic vein was 5.7 ± 2.3, 4.2 ± 1.9, and 2.7 ± 1.4, respectively with subjective scores of 4.3 ± 0.3, 3.8 ± 0.4, and 3.2 ± 0.3, respectively ( < 0.001).
Optimal monochromatic images can lead to improvement in the imaging parameters and optimization of the image quality of the common hepatic artery, hepatic portal vein and hepatic vein compared with conventional mixed kV and with 70 keV datasets.
研究能谱 CT 最佳单能量图像在提高肝脏血管成像质量方面的效率。
回顾性分析 35 例腹部 CT 血管造影的影像学资料。采用混合能量(质量检查,QC)、70keV 和最佳单能量模式重建肝动脉、门静脉和肝静脉。分析比较 CT 值、图像噪声(IN)、对比噪声比(CNR)、信噪比(SNR)和主观定性分析等参数。
评估肝总动脉、门静脉和肝静脉的最佳单能量值在 49keV 到 53keV 之间,平均为 51keV。最佳单能量、70keV 和 QC 图像在肝血管 CT 值、IN、CNR、SNR 和主观定性评分方面均有统计学差异(<0.001)。肝总动脉的 CNR 在最佳单能量组、70keV 组和 QC 组分别为 24.6±10.9、18.1±8.3 和 11.6±4.6(<0.001),主观评分分别为 4.7±0.2、4.0±0.3 和 3.6±0.4(<0.001)。门静脉 CNR 分别为 6.9±2.7、4.3±1.9 和 3.0±2.1(<0.001),主观评分分别为 4.5±0.3、3.9±0.4 和 3.3±0.3(<0.001)。肝静脉 CNR 分别为 5.7±2.3、4.2±1.9 和 2.7±1.4,主观评分分别为 4.3±0.3、3.8±0.4 和 3.2±0.3(<0.001)。
与常规混合 keV 图像和 70keV 数据集相比,最佳单能量图像可改善肝总动脉、门静脉和肝静脉的成像参数和图像质量。