Comparison of cross sectional optical coherence tomography images of elevated optic nerve heads across acquisition devices and scan protocols.

作者信息

Patel Megh D, Khushzad Fareshta, Moss Heather E

机构信息

1Department of Ophthalmology, Stanford University, 2370 Watson Court, suite 200, Palo Alto, CA 94303 USA.

2Department of Neurology and Neurological Sciences, Stanford University, 2370 Watson Court, suite 200, Palo Alto, CA 94303 USA.

出版信息

Eye Vis (Lond). 2018 Jul 14;5:17. doi: 10.1186/s40662-018-0112-3. eCollection 2018.

Abstract

BACKGROUND

Optic nerve head measurements extracted from optical coherence tomography (OCT) show promise for monitoring clinical conditions with elevated optic nerve heads. The aim of this study is to compare reliability within and between raters and between image acquisition devices of optic nerve measurements derived from OCT scans in eyes with varying degrees of optic nerve elevation.

METHODS

Wide angle line scans and narrow angle radial scans through optic nerve heads were obtained using three spectral domain(SD) OCT devices on 5 subjects (6 swollen optic nerves, 4 normal optic nerves). Three raters independently semi-manually segmented the internal limiting membrane(ILM) and Bruch's membrane(BM) on each scan using customized software. One rater segmented each scan twice. Segmentations were qualitatively and quantitatively compared. Inter-rater, intra-rater and inter-device reliability was assessed for the optic nerve cross sectional area calculated from the ILM and BM segmentations using intraclass correlation coefficients and graphical comparison.

RESULTS

Line scans from all devices were qualitatively similar. Radial scans for which frame rate could not be adjusted were of lower quality. Intra-rater reliability for segmentation and optic nerve cross sectional area was better than inter-rater reliability, which was better than inter-device reliability, though all ICC exceeded 0.95. Reliability was not impacted by the degree of optic nerve elevation.

CONCLUSIONS

SD-OCT devices acquired similar quality scans of the optic nerve head, with choice of scan protocol affecting the quality. For image derived markers, variability between devices was greater than that attributable to inter and intra-rater differences.

摘要

背景

从光学相干断层扫描(OCT)中提取的视神经乳头测量值有望用于监测视神经乳头升高的临床情况。本研究的目的是比较不同程度视神经升高的眼睛中,评估者内部和之间以及图像采集设备之间,源自OCT扫描的视神经测量值的可靠性。

方法

使用三台光谱域(SD)OCT设备对5名受试者(6只肿胀视神经、4只正常视神经)进行穿过视神经乳头的广角线扫描和窄角径向扫描。三名评估者使用定制软件在每次扫描上独立地半自动分割内界膜(ILM)和布鲁赫膜(BM)。一名评估者对每次扫描进行两次分割。对分割结果进行定性和定量比较。使用组内相关系数和图形比较,评估从ILM和BM分割计算出的视神经横截面积的评估者间、评估者内和设备间可靠性。

结果

所有设备的线扫描在质量上相似。帧率无法调整的径向扫描质量较低。分割和视神经横截面积的评估者内可靠性优于评估者间可靠性,评估者间可靠性优于设备间可靠性,不过所有组内相关系数均超过0.95。可靠性不受视神经升高程度的影响。

结论

SD-OCT设备对视神经乳头的扫描质量相似,扫描协议的选择会影响质量。对于图像衍生标记物,设备间的变异性大于评估者间和评估者内差异所致的变异性。