Macleod Kirsty, Katz Laurence B, Cameron Hilary
1 LifeScan Scotland Ltd, Inverness, UK.
2 LifeScan, Inc, West Chester, PA, USA.
J Diabetes Sci Technol. 2019 May;13(3):546-552. doi: 10.1177/1932296818790228. Epub 2018 Aug 7.
Anecdotal blood glucose assessments conducted by health care professionals (HCPs) in the field have highlighted differences in results when methodology used is not according to best practices for measuring blood glucose. This study assessed the impact on accuracy of blood glucose measurements when methodology deviates from the recommended study design and recommended reference instrument.
Adults with type 1 or type 2 diabetes provided capillary and venous blood samples for accuracy assessments using OneTouch® Verio® (Verio) and OneTouch® Ultra 2® (Ultra) blood glucose meters (BGM) and two different reference instruments.
Increases in mean bias were observed when comparing capillary to venous samples tested on the BGMs and the recommended reference instrument. Mean bias was even greater when a hospital blood glucose analyzer was used to measure venous plasma glucose. Increases in mean bias observed for Ultra BGM when testing venous blood on the meter compared to the recommended reference instrument was likely due to the interfering effects of low oxygen levels in the venous blood sample. Conversely, Verio meters, which are insensitive to low oxygen levels, showed little difference from baseline when testing venous blood on the meter compared to results from the same venous sample measured on a reference instrument.
Deviations from the best practice study design of comparing capillary blood glucose results tested on the blood glucose meter with the manufacturer's stated reference instrument will affect accuracy of blood glucose measurements.