Suppr超能文献

溯源至一级参考测量程序(ID-LCMS);验证医院血糖仪临床准确性和安全性的关键步骤。

Traceability to a primary reference measurement procedure (ID-LCMS); A key step in validating the clinical accuracy and safety of hospital blood glucose monitoring systems.

机构信息

'Department of Endocrinology and Metabolism, Shanghai Clinical Center for Diabetes, Shanghai Diabetes Institute, Shanghai Key Laboratory of Diabetes Mellitus, Shanghai Jiao Tong University Affiliated Sixth People's Hospital, Shanghai 200233, China.

School of Environmental Science and Engineering, Shanghai Jiao Tong University, Shanghai 200240, China.

出版信息

Clin Chim Acta. 2018 Nov;486:275-281. doi: 10.1016/j.cca.2018.08.025. Epub 2018 Aug 18.

Abstract

OBJECTIVE

A key step in the evaluation of the accuracy of blood glucose monitoring systems (BGMS) is using a comparator method aligned to a high order definitive reference method. We describe how we achieved traceability to an isotope dilution liquid chromatography mass spectrometry (ID-LCMS) method. We used ID-LCMS to evaluate the accuracy and specificity of two hospital BGMS used in China.

METHOD

ID-LCMS was used to verify the calibration alignment of the laboratory plasma hexokinase reference method using NIST standard reference material and clinical samples. The ID-LCMS aligned hexokinase method was used to evaluate the clinical accuracy of two BGMS in hospitalized patients. System accuracy was evaluated using Chinese consensus guidelines. BGMS accuracy was also assessed with interference factors known to be present in critically ill patients' blood.

RESULTS

The laboratory plasma hexokinase reference method was shown to calibrate closely with ID-LCMS. Two BGMS demonstrated good correlation with this reference method. Only one BGMS met the Chinese guidelines. The interference factors didn't influence this BGMS but adversely affected the clinical accuracy of the other.

CONCLUSIONS

We advocate that our IDMS calibration alignment approach for ensuring the accuracy of the glucose reference method should be adopted in evaluations assessing the accuracy of blood glucose monitoring systems.

摘要

目的

评估血糖监测系统(BGMS)准确性的关键步骤是使用与高级明确参考方法一致的比较器方法。我们描述了如何实现与同位素稀释液相色谱质谱法(ID-LCMS)方法的可追溯性。我们使用 ID-LCMS 评估了在中国使用的两种医院 BGMS 的准确性和特异性。

方法

使用 ID-LCMS 通过 NIST 标准参考物质和临床样本验证实验室血浆己糖激酶参考方法的校准对齐。使用 ID-LCMS 对齐的己糖激酶方法评估住院患者中两种 BGMS 的临床准确性。使用中国共识指南评估系统准确性。还评估了已知存在于危重症患者血液中的干扰因素对 BGMS 准确性的影响。

结果

实验室血浆己糖激酶参考方法与 ID-LCMS 紧密校准。两种 BGMS 与该参考方法具有良好的相关性。只有一种 BGMS 符合中国指南。干扰因素并未影响该 BGMS,但对另一种 BGMS 的临床准确性产生了不利影响。

结论

我们主张采用我们的 IDMS 校准对齐方法来确保葡萄糖参考方法的准确性,以评估血糖监测系统的准确性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验