Angelides Kimon, Matsunami Risë K, Engler David A
Livongo Health, Inc, Palo Alto, CA, USA
Proteomics Programmatic Core Laboratory, Houston Methodist Hospital Research Institute, Houston, TX, USA.
J Diabetes Sci Technol. 2015 May 22;9(5):1061-70. doi: 10.1177/1932296815586016.
We evaluated the accuracy, precision, and linearity of the In Touch blood glucose monitoring system (BGMS), a new color touch screen and cellular-enabled blood glucose meter, using a new rapid, highly precise and accurate (13)C6 isotope-dilution liquid chromatography-mass spectrometry method (IDLC-MS).
Blood glucose measurements from the In Touch BGMS were referenced to a validated UPLC-MRM standard reference measurement procedure previously shown to be highly accurate and precise. Readings from the In Touch BGMS were taken over the blood glucose range of 24-640 mg/dL using 12 concentrations of blood glucose. Ten In Touch BGMS and 3 lots of test strips were used with 10 replicates at each concentration. A lay user study was also performed to assess the ease of use.
At blood glucose concentrations <75 mg/dL 100% of the measurements are within ±8 mg/dL from the true reference standard; at blood glucose levels >75 mg/dL 100% of the measurements are within ±15% of the true reference standard. 100% of the results are within category A of the consensus grid. Within-run precision show CV < 3.72% between 24-50 mg/dL and CV<2.22% between 500 and 600 mg/dL. The results show that the In Touch meter exceeds the minimum criteria of both the ISO 15197:2003 and ISO 15197:2013 standards. The results from a user panel show that 100% of the respondents reported that the color touch screen, with its graphic user interface (GUI), is well labeled and easy to navigate.
To our knowledge this is the first touch screen glucose meter and the first study where accuracy of a new BGMS has been measured against a true primary reference standard, namely IDLC-MS.
我们使用一种新的快速、高精度且准确的(13)C6同位素稀释液相色谱 - 质谱法(IDLC - MS),评估了新型彩色触摸屏且具备蜂窝功能的血糖仪In Touch血糖监测系统(BGMS)的准确性、精密度和线性。
将In Touch BGMS的血糖测量结果与先前已证明具有高准确性和精密度的经过验证的超高效液相色谱 - 多反应监测(UPLC - MRM)标准参考测量程序进行比对。使用12种血糖浓度,在24 - 640 mg/dL的血糖范围内获取In Touch BGMS的读数。使用了10台In Touch BGMS和3批测试条,每种浓度重复测量10次。还进行了一项普通用户研究以评估其易用性。
在血糖浓度<75 mg/dL时,100%的测量值与真实参考标准的偏差在±8 mg/dL以内;在血糖水平>75 mg/dL时,100%的测量值在真实参考标准的±15%以内。100%的结果处于共识网格的A类。批内精密度在24 - 50 mg/dL之间CV<3.72%,在500 - 600 mg/dL之间CV<2.22%。结果表明In Touch血糖仪超过了ISO 15197:2003和ISO 15197:2013标准的最低标准。用户小组的结果显示,100%的受访者表示带有图形用户界面(GUI)的彩色触摸屏标签清晰且易于操作。
据我们所知,这是首个触摸屏血糖仪,也是首次针对新型BGMS与真正的一级参考标准即IDLC - MS进行准确性测量的研究。