OBJECTIVE

Acoustic radiation force impulse (ARFI) imaging is a recent ultrasound elastography technique; consequently, its efficacy is not fully known. In this study, we compared ARFI imaging with conventional strain elastography (SE) and shear wave velocities (SWVs) to evaluate the utility of ARFI imaging for diagnosing thyroid nodules.

SUBJECTS AND METHODS

In this study we examined 233 thyroid nodules (183 benign nodules and 50 malignant nodules) isolated from human patients. The nodules were evaluated with SE and ARFI imaging, and SWVs of the nodules were simultaneously measured. ARFI images were classified using a four-point score based on grayscale intensity of the images. The sensitivity, specificity, and diagnostic accuracy were compared between SE and ARFI imaging. Finally, SWVs for each score of SE and ARFI imaging were compared.

RESULTS

The new scoring system for ARFI imaging can be divided into four virtual touch imaging (VTI) scores. Nodules with a VTI score of 3 or 4 as determined by ARFI imaging were determined to be malignant. The sensitivity, specificity, and diagnostic accuracy, respectively, were 63.2%, 66.3%, and 65.6% for SE, compared with 80.0%, 86.3%, and 85.0% for ARFI imaging. The median SWVs of the nodules were 1.57 m/s, 1.73 m/s, 1.88 m/s, and 2.09 m/s for VTI scores of 1, 2, 3, and 4, respectively. The SWVs of VTI scores 3 and 4 were significantly higher than those of VTI scores 1 and 2.

CONCLUSIONS

The diagnostic accuracy of ARFI imaging for differentiating malignant thyroid nodules was higher than that of SE. The VTI scores of the nodules accurately reflected their SWVs.

LEVEL OF EVIDENCE

摘要

目的

声辐射力脉冲(ARFI)成像技术是一种新型超声弹性成像技术,其有效性尚未完全明确。本研究通过比较ARFI成像与传统应变弹性成像(SE)及剪切波速度(SWV),评估ARFI成像在甲状腺结节诊断中的应用价值。

对象与方法

本研究纳入233例来自人类患者的甲状腺结节(183例良性结节和50例恶性结节)。采用SE和ARFI成像对结节进行评估,并同步测量结节的SWV。基于图像灰度强度,采用四点评分法对ARFI图像进行分类。比较SE和ARFI成像的敏感性、特异性及诊断准确性。最后,比较SE和ARFI成像各评分对应的SWV。

结果

ARFI成像新评分系统可分为四个虚拟触诊成像(VTI)评分。ARFI成像VTI评分为3或4的结节判定为恶性。SE的敏感性、特异性及诊断准确性分别为63.2%、66.3%和65.6%,而ARFI成像分别为80.0%、86.3%和85.0%。VTI评分1、2、3和4对应的结节SWV中位数分别为1.57 m/s、1.73 m/s、1.88 m/s和2.09 m/s。VTI评分3和4的SWV显著高于VTI评分1和2。

结论

ARFI成像鉴别甲状腺恶性结节的诊断准确性高于SE。结节的VTI评分准确反映了其SWV。

证据等级

4级。