Biosens Bioelectron. 2018 Dec 15;121:41-46. doi: 10.1016/j.bios.2018.08.067. Epub 2018 Aug 30.
Prostate specific antigen (PSA) is a widely used marker for the diagnosis of prostate cancer, and the increasing attention has been attracted on the development of rapid assay using biosensing technology. However, it remains challenging for the sensitive and selective detection of PSA in clinical samples. Here, we report a label-free microfluidic paper-based analytical device for highly sensitive electrochemical detection of PSA. The paper device was fabricated with wax printing to generate hydrophobic and hydrophilic layers for the construction of microfluidic channel, followed by screen-printing of three electrodes including working, counter and reference electrode. Gold nanoparticles (AuNPs)/reduced graphene oxide (rGO)/thionine (THI) nano composites were synthesized and characterized, which were coated onto working electrodes for the immobilization of DNA aptamer probe. THI servers as the electrochemical mediator to transduce the biological recognition between DNA aptamer and PSA, and the excellent conductivity of AuNPs and rGO also play a significant role of electron transfer, leading to a sensitive detection for PSA, able to detect PSA as low as 10 pg mL, with a linear range from 0.05 to 200 ng mL. We demonstrated that our electrochemical sensor for the detection of clinical serum samples, indicating that our sensor would provide a new platform for low cost, sensitive and point-of-care diagnosis of prostate cancer.
前列腺特异性抗原(PSA)是诊断前列腺癌的常用标志物,基于生物传感技术的快速检测方法受到了越来越多的关注。然而,在临床样本中实现 PSA 的灵敏和选择性检测仍然具有挑战性。在此,我们报告了一种用于 PSA 高灵敏电化学检测的无标记微流控纸基分析器件。该纸器件采用蜡印技术来生成用于构建微流道的疏水和亲水层,然后通过丝网印刷技术制备工作电极、对电极和参比电极。合成并表征了金纳米粒子(AuNPs)/还原氧化石墨烯(rGO)/硫堇(THI)纳米复合材料,并将其涂覆到工作电极上以固定 DNA 适体探针。THI 用作电化学介体,以转导 DNA 适体与 PSA 之间的生物识别,AuNPs 和 rGO 的优异导电性也在电子转移中发挥重要作用,从而实现对 PSA 的灵敏检测,能够检测到低至 10pg/mL 的 PSA,线性范围为 0.05 至 200ng/mL。我们证明了我们的电化学传感器可用于检测临床血清样本,表明我们的传感器将为低成本、灵敏和即时诊断前列腺癌提供新平台。