Lyon University, CNRS, Inserm, CREATIS UMR5220, France. CNRS/IN2P3 and Lyon 1 University, UMR 5822, Villeurbanne, France.
Phys Med Biol. 2018 Oct 17;63(20):205016. doi: 10.1088/1361-6560/aae043.
Proton transmission imaging uses protons with high enough energy to fully traverse the phantom/patient and to be captured in a suitable detector placed behind it. The measured residual energy or residual range provide a direct estimate of the water equivalent thickness (WET) of the image volume. Requirements for proton imaging to be exploitable in clinical practice include: sufficient WET accuracy and integrability into the treatment room and the clinical workflow, as well as an acceptably low dose to the patient and a sufficient spatial resolution. In this work, we report on experiments performed at the Institut Curie-Proton therapy center in Orsay (IC-CPO), France, using a commercial range telescope commonly employed for quality assurance measurements. The purpose was to keep the experimental set-up as simple as possible and to achieve nonetheless high WET accuracy radiographies by developing and applying dedicated post processing methods. We explain these methods in detail and discuss their performance. We assess the WET accuracy based on two different reference phantoms: a CIRS electron density phantom with tissue equivalent inserts and a homogeneous step phantom. We find an agreement between the measured and the reference WET values of 0.2-0.5 mm. The lowest investigated dose was 10 mGy per acquisition. It could be lowered by modifying the irradiation plan and lowering the beam current, though the latter would impose slight optimisations of the detector hardware. Our work suggests that proton radiographies with good WET accuracy can be obtained with a reasonable experimental effort that would facilitate integration into clinical routine.
质子透射成像使用能量足够高的质子,使其能够完全穿透体模/患者,并被放置在其后面的合适探测器捕获。测量的剩余能量或剩余射程提供了图像体积的水等效厚度 (WET) 的直接估计。质子成像要在临床实践中得到应用,需要满足以下要求:足够的 WET 精度和可集成性,以及可接受的患者剂量和足够的空间分辨率。在这项工作中,我们报告了在法国奥尔赛的居里研究所质子治疗中心 (IC-CPO) 进行的实验,使用了常用于质量保证测量的商业射程望远镜。目的是尽可能保持实验设置简单,并通过开发和应用专用的后处理方法来实现高 WET 精度射线照相。我们详细解释了这些方法,并讨论了它们的性能。我们根据两个不同的参考体模评估了 WET 精度:一个具有组织等效插件的 CIRS 电子密度体模和一个均匀台阶体模。我们发现测量的和参考的 WET 值之间的差异在 0.2-0.5 毫米之间。最低研究剂量为每次采集 10 毫戈瑞。通过修改照射计划和降低束流,可以降低剂量,但后者将对探测器硬件进行轻微优化。我们的工作表明,通过合理的实验努力,可以获得具有良好 WET 精度的质子射线照相,这将有助于将其集成到临床常规中。