VersaTrek 528血培养系统在中国一家三级医院的临床性能评估

Clinical Performance Evaluation of VersaTrek 528 Blood Culture System in a Chinese Tertiary Hospital.

作者信息

Yue Pinli, Zhou Menglan, Kudinha Timothy, Xie Xiuli, Du Juan, Song Hongmei, Zhang Lintao, Ma Xiaojun, Weng Li, Chai Wenzhao, Zhu Huadong, Yang Qiwen, Xu Ying-Chun

机构信息

Department of Clinical Laboratory, Peking Union Medical College Hospital, Peking Union Medical College, Chinese Academy of Medical Sciences, Beijing, China.

Beijing Key Laboratory for Mechanisms Research and Precision Diagnosis of Invasive Fungal Diseases, Beijing, China.

出版信息

Front Microbiol. 2018 Aug 28;9:2027. doi: 10.3389/fmicb.2018.02027. eCollection 2018.

Abstract

The aim of this study was to evaluate the clinical performance of VersaTrek 528 compared to BACTEC FX 400 blood culture (BC) systems. Simulated and clinically obtained BCs were used in the study. Confirmed bacterial species ( = 78), including 43 Gram-positives, 30 Gram-negatives, and 5 strains, were each inoculated into BC bottles. Clinically obtained BCs were subdivided into two groups, A and B. In group A were 72 BC sets (pair: aerobic and anaerobic) in which a set inoculated with 5 ml blood was processed in the VersaTrek BC system, whilst the one inoculated with 10 ml blood was processed in the FX BC system. In group B, 76 BC sets (pairs) corresponding to 152 VersaTrek bottles and 152 FX bottles were inoculated with the same volume (10 ml) of blood, and processed in each system. In the simulated BC study, 90% (63/70) of the VersaTrek aerobic bottles were positive, which was higher than that of FX 400 (59/70, 84%), but was not statistically significant ( = 0.423). In contrast, FX 400 anaerobic bottles had a higher positive rate than the other BC system (84 vs. 77%), although it was statistically insignificant ( = 0.267). Time to detection of organisms in the two BCs was comparable for both aerobic ( = 0.131) and anaerobic bottles ( = 0.104). In clinical BCs of group A, FX BC system had slightly higher positive rates for both aerobic (11.1 vs. 9.7%, = 0.312) and anaerobic (8.3 vs. 6.9%, = 0.375) bottles. However, the difference was not statistically significant. In group B, VersaTrek aerobic bottles had a higher positive rate compared to the other BC system (10.5 vs. 5.2%, = 0.063). In terms of positive rates of sub-studies A and B, VersaTrek and FX BC systems were comparable. There was no significant difference between the two BC systems in the detection of bacteria and fungi in simulated BCs. In clinical BCs, the performance of the VersaTrek BC system, with inoculation of 5 or 10 ml patient's blood, was comparable to the FX system with inoculation of 10 ml patient's blood.

摘要

本研究的目的是评估VersaTrek 528与BACTEC FX 400血培养(BC)系统相比的临床性能。研究中使用了模拟和临床获取的血培养样本。将确诊的细菌种类(n = 78),包括43株革兰氏阳性菌、30株革兰氏阴性菌和5株其他菌株,分别接种到血培养瓶中。临床获取的血培养样本被分为A、B两组。A组有72套血培养样本(配对:需氧和厌氧),其中一套接种5 ml血液的样本在VersaTrek血培养系统中处理,而接种10 ml血液的样本在FX血培养系统中处理。B组,76套血培养样本(配对),对应152个VersaTrek瓶和152个FX瓶,接种相同体积(10 ml)的血液,并在每个系统中进行处理。在模拟血培养研究中,VersaTrek需氧瓶的阳性率为90%(63/70),高于FX 400(59/70,84%),但差异无统计学意义(P = 0.423)。相比之下,FX 400厌氧瓶的阳性率高于另一个血培养系统(84%对77%),尽管差异无统计学意义(P = 0.267)。两种血培养系统中需氧瓶(P = 0.131)和厌氧瓶(P = 0.104)检测到微生物的时间相当。在A组的临床血培养样本中,FX血培养系统在需氧瓶(11.1%对9.7%,P = 0.312)和厌氧瓶(8.3%对6.9%,P = 0.375)中的阳性率略高。然而,差异无统计学意义。在B组中,VersaTrek需氧瓶的阳性率高于另一个血培养系统(10.5%对5.2%,P = 0.063)。就A、B子研究的阳性率而言,VersaTrek和FX血培养系统相当。在模拟血培养中,两种血培养系统在检测细菌和真菌方面无显著差异。在临床血培养中,接种5或10 ml患者血液的VersaTrek血培养系统的性能与接种10 ml患者血液的FX系统相当。