Park Joonhong, Han Seounghwan, Shin Soyoung
Clin Lab. 2017 Jan 1;63(1):39-46. doi: 10.7754/Clin.Lab.2016.160502.
We evaluated growth performance of the BacT/ALERT VIRTUO and the BACTEC FX blood culture (BC) systems based on recovery rate and time to detect microorganisms under simulated bloodstream infection conditions.
Three expected concentrations of 125, 30, and 5 CFU/mL of eighteen reference microorganisms were diluted with 5 mL blood for adult aerobic/anaerobic culture and with 3 mL for pediatric aerobic culture and inoculated, and each of three BC bottles were loaded into the two BC systems.
A total of 405 comparative bottles were analyzed in this study. The percent agreement between the BacT/ ALERT VIRTUO and BACTEC FX systems after comparing the positive results was 96.8% (392/405). The BacT/ ALERT VIRTUO and BACTEC FX systems yielded 100% positive signals at 125 and 30 CFU/mL, whereas there were 5.2% (7/135) and 8.1% (11/135) false-negative results at 5 CFU/mL, respectively. No statistical difference in recovery rate based on the overall concentration of microorganisms was observed between the two BC systems (p = 0.329). No statistical difference in time to detection (TTD) was observed between the two BC systems based on the overall concentration of microorganisms (p = 0.067). Inter-instrument comparisons of TTD resulted in a good correlation (r = 0.947) for all inoculated bottles.
The BacT/ALERT VIRTUO showed similar recovery rate and TTD of microorganisms, compared to BACTEC FX BC system. Due to its versatility to accommodate a higher workload through automated loading and unloading of BC bottles with faster detection of pathogens from bloodstream, the BacT/ALERT VIRTUO system may establish itself as a useful alternative.
我们基于在模拟血流感染条件下微生物的回收率和检测时间,评估了BacT/ALERT VIRTUO和BACTEC FX血培养(BC)系统的生长性能。
将18种参考微生物的三种预期浓度125、30和5 CFU/mL用5 mL血液稀释用于成人需氧/厌氧培养,用3 mL血液稀释用于儿科需氧培养,然后进行接种,并将三个BC瓶分别装入这两种BC系统。
本研究共分析了405对比较瓶。比较阳性结果后,BacT/ALERT VIRTUO和BACTEC FX系统之间的一致率为96.8%(392/405)。BacT/ALERT VIRTUO和BACTEC FX系统在125和30 CFU/mL时产生100%的阳性信号,而在5 CFU/mL时分别有5.2%(7/135)和8.1%(11/135)的假阴性结果。两种BC系统之间基于微生物总体浓度的回收率无统计学差异(p = 0.329)。基于微生物总体浓度,两种BC系统之间的检测时间(TTD)无统计学差异(p = 0.067)。所有接种瓶的TTD仪器间比较显示出良好的相关性(r = 0.947)。
与BACTEC FX BC系统相比,BacT/ALERT VIRTUO显示出相似的微生物回收率和TTD。由于其通过自动装卸BC瓶来适应更高工作量的多功能性,以及能更快地从血流中检测病原体,BacT/ALERT VIRTUO系统可能成为一种有用的替代方案。